Measurements included beam quality, output and radiation field size and uniformity.. Conclusions: The dosimetric characterisation of the Papillon 50 was validated by the audit measuremen
Trang 1Original Research Article
National audit of a system for rectal contact brachytherapy
Laia Humbert-Vidana,⇑, Thorsten Sanderb, David J Eatonc, Catharine H Clarkb,c,d
a
Department of Medical Physics, St Thomas’ Hospital, London, UK
b Radiation Dosimetry Group, National Physical Laboratory, Teddington, Middlesex, UK
c
National Radiotherapy Trials QA (RTTQA) Group, Mount Vernon Hospital, Northwood, Middlesex, UK
d
Department of Medical Physics, Royal Surrey County Hospital, Guildford, Surrey, UK
a r t i c l e i n f o
Article history:
Received 12 July 2016
Received in revised form 4 December 2016
Accepted 4 December 2016
Keywords:
Contact brachytherapy
Electronic brachytherapy
Audit
a b s t r a c t Background and purpose: Contact brachytherapy is used for the treatment of early rectal cancer An over-view of the current status of quality assurance of the rectal contact brachytherapy systems in the UK, based on a national audit, was undertaken in order to assist users in optimising their own practices Material and methods: Four UK centres using the Papillon 50 contact brachytherapy system were audited Measurements included beam quality, output and radiation field size and uniformity Test frequencies and tolerances were reviewed and compared to both existing recommendations and published reviews
on other kV and electronic brachytherapy systems External validation of dosimetric measurements was provided by the National Physical Laboratory
Results: The maximum host/audit discrepancy in beam quality determination was 6.5%; this resulted in absorbed dose variations of 0.2% The host/audit agreement in absorbed dose determination was within 2.2% The median of the radiation field uniformity measurements was 2.7% and the host/audit agreement
in field size was within 1 mm Test tolerances and frequencies were within the national recommenda-tions for kV units
Conclusions: The dosimetric characterisation of the Papillon 50 was validated by the audit measurements for all participating centres, thus providing reassurance that the implementation had been performed within the standards stated in previously published audit work and recommendations for kV and elec-tronic brachytherapy units However, optimised and standardised quality assurance testing could be achieved by reducing some methodological differences observed
Ó 2017 The Authors Published by Elsevier Ireland Ltd on behalf of European Society of Radiotherapy &
licenses/by-nc-nd/4.0/)
1 Introduction
Regular dosimetric intercomparison has been undertaken in the
UK for the past 30 years[1] During this time audit groups in the
UK have been developing and improving audit programmes with
the aim of reducing the practice variability between radiotherapy
departments [1,2]and maintaining quality standards across the
country An independent audit is especially useful when
imple-menting new techniques for which commissioning and quality
assurance guidelines or recommendations are not yet in place In
2015 the National Institute of Health and Care Excellence (NICE)
issued guidance on safety and efficacy of the rectal contact
brachytherapy technique from a clinical perspective[3] However,
as far as we know, there is currently no guidance on equipment quality assurance testing Electronic brachytherapy devices repre-sent a 15% of the kV treatment units in the UK[4] The aim of this audit was to perform a dosimetric intercomparison of the different centres and to provide an overview of the current practice in qual-ity assurance of the systems used for rectal contact brachytherapy
in the UK in order to assist current and future users to optimise their own practices as well as to establish a methodology and tol-erances for future audits
A contact brachytherapy system was released in 2008 for the treatment of early rectal cancer It is used for conservative treatment
as an alternative to radical surgery for patients at a higher anaes-thetic risk or who are willing to accept a higher recurrence risk in order to avoid a permanent colostomy [5] Contact radiotherapy can also be used as adjuvant radiotherapy to local resection, with
50 Gy usually delivered in 3 fractions, or as a boost to external beam radiotherapy, with 90–110 Gy delivered in 3 fractions[6]
http://dx.doi.org/10.1016/j.phro.2016.12.001
2405-6316/Ó 2017 The Authors Published by Elsevier Ireland Ltd on behalf of European Society of Radiotherapy & Oncology.
⇑ Corresponding author at: Department of Medical Physics, St Thomas’ Hospital,
Guy’s and St Thomas’ NHS Foundation Trust, Westminster Bridge Road, London SE1
7EH, UK.
E-mail addresses: laia.humbert-vidan@nhs.net (L Humbert-Vidan), thorsten.
sander@npl.co.uk (T Sander), davideaton@nhs.net (D.J Eaton), catharine.clark@nhs.
net (C.H Clark).
Contents lists available atScienceDirect
Physics and Imaging in Radiation Oncology
j o u r n a l h o m e p a g e : w w w e l s e v i e r c o m / l o c a t e / p h r o
Trang 22 Materials and methods
Four centres participated in the audit, with Papillon 50 contact
brachytherapy systems (Ariane Medical Systems, Ltd, Derby, UK)
commissioned between 2009 and 2014 and a workload of 1–30
patients per month A ‘single auditor’ approach with on-site visits
was taken as a more consistent and simplified analysis
methodol-ogy was easier to achieve with centrally organised audits[7–9]
Treatment with the Papillon 50 is delivered with a hand-guided
X-ray tube that produces a 50 kVp and approximately 2.7 mA beam
with dose rates as high as 15 Gy/min Electrons are accelerated
towards a rhenium transmission target and photons are produced
isotropically The focus-to-surface distance (FSD) of the applicators
(29, 32 and 38 mm) varies with applicator diameter (22, 25 and
30 mm, respectively) in order to achieve a collimated beam with
a fixed opening of 45°[6,10,11]
The audit measurements included beam quality, radiation
out-put, and radiation field size and uniformity A comparison between
host and audit measurements was made, with a discussion of the
significance of the differences observed The National Physical
Laboratory (NPL, Teddington) provided external validation of the
procedures during the visit to the first audited centre[8] Most of
the dosimetry equipment used was provided by NPL, thus direct
traceability for all audit results to the national standard was
ensured In addition, constancy checks using a strontium check
source were carried out on the ionisation chamber by NPL before
the first visit and after the last visit of the audit A review was
car-ried out on the quality assurance programme documentation
pro-vided by all centres[12]; this included tolerances and frequencies
of tests following their respective ISO 9000 Quality Systems A
comparison was made (Table 1) to IPEM 81 recommendations
[13]and to a recent review on electronic brachytherapy[6]
2.1 Beam quality (HVL)
Peak tube potential and first half-value layer (HVL1) are the
rec-ommended beam quality specifiers for very low energy X-ray
beams, such as that produced by the Papillon 50 unit The IPEMB
code of practice (CoP) for the determination of absorbed dose for
X-rays below 300 kV generating potential[14]recommends scatter free
and narrow beam geometry for the HVL measurement Each centre
had designed their own custom-built HVL jig (seeTable 1andFig 1
in Supplementary material) to achieve such measurement
condi-tions; the audit HVL jig was borrowed from centre C A PTW type
23342 0.02 cm3soft X-ray thin-window secondary standard
paral-lel plate ionisation chamber calibrated in terms of air kerma and a
calibrated Scanditronix Wellhofer type Dose 1 electrometer were
used All centres used the same ionisation chamber model and all
the equipment was calibrated, traceable to the national standard
Temperature and pressure were measured with a Digitron
hand-held thermometer type 2024T and a Greisinger electronic
barome-ter model GTD 1100, respectively Six 99.999% purity aluminium
filters were customised for this audit and their thicknesses
mea-sured at NPL with a calibrated coordinate measuring machine;
the standard deviation of the thickness measurements ranged from
approximately 0.002 mmAl to 0.003 mmAl for the thinnest
(0.0571 mmAl) and thickest (1.039 mmAl) filters, respectively
The audited centres used their own Al filters for their
measure-ments Exposures of 500 MU were performed with increasing
levels of attenuation using the aluminium filters Repeat readings
were corrected for temperature and pressure and the mean value
plotted against the total thickness of added aluminium The HVL
value was derived from a second-degree polynomial fit and
compared to the host HVL value The effect of host-audit HVL
dis-crepancies on the determination of absorbed dose to water was
assessed
a /m
Dw
Trang 32.2 Radiation output
All Papillon units were calibrated to deliver 30 Gy for a
3000 MU exposure The absorbed dose to water at the surface of
a full-scatter water-equivalent phantom (Dw) was determined by
applying the formalism from the IPEM very low energy
(0.035–1.0 mmAl) CoP[14,15] The calibration coefficients for the
chamber and electrometer were obtained from the respective
cal-ibration certificates by NPL, with a quoted uncertainty of 1.2%
(k = 2) for the chamber The host’s quoted HVL value was used to
approximate the relevant mass energy coefficient ratio in air,
near-est values on the reference tables[14–16] The audit measured HVL
value was not used to calculate the radiation output in order to
exclude any uncertainties or differences due to the HVL jig design
By using the host’s HVL value, the only differences in measurement
setup/equipment were those related exclusively to the output
measurement NPL use a radiation field with circular
cross-section and 53 mm diameter whereas this audit used the 30 mm
applicator; Perrin et al (2001)[17]have shown that there is no
change in kchwith field size for the PTW 23442 chamber below a
40 mm diameter field size They also state that the kch is not
expected to change with a variation in FSD
A poly methyl methacrylate (PMMA, Perspex) output jig was
originally designed by centre A (Fig 1) Based on that design,
Ariane produced their own polyoxymethylene (POM, Delrin) jig
for output and film measurements, which is now provided with
the Papillon 50 unit (Fig 2 in Supplementary material) For
conve-nience, audit measurements were carried out in this output jig
instead of a full-scatter phantom A comparison was made by
cen-tre A between absorbed doses measured in a full scatter phantom
and in the output jig to assess the uncertainty introduced by such
non full-scatter conditions
Centres measured the temperature differently: be it room
temperature, chamber temperature or output jig temperature
The latter was measured by centre B by means of a custom-built
POM jig, which was fitted into the chamber slot within the output
jig The effect of these differences on the resulting radiation output
was investigated
Overall uncertainty in HVL (1.7%) and Dw(2.1%) determination
was derived from the root-mean-square of the following estimated
uncertainties[12]: measurement reproducibility (1.0% and 0.3% for
HVL and Dw, respectively), measurements of temperature (0.2%)
and pressure (0.1%), electrometer precision (0.04%), Al filter
thick-ness (1.7%), effect of HVL set up variation on determination of Dw
(0.2%) and electrometer (0.2%) and chamber (1.2%) calibrations
2.3 Radiation field size and uniformity
For each applicator size two pieces of RTQA Gafchromic film
were exposed to 1 Gy and 2 Gy, respectively; dose linearity was
assessed with charge readings at increasing MU levels to
deter-mine the required MU for doubling the dose level for the second film exposure Films were scanned one week after their exposure using an Epsom type 11000 Pro flatbed scanner (professional mode, reflective scan, 48-bit colour, 96 dpi) Most centres used the jig with the approximately 2 cm thick backscatter block flush against the film rather than using the gap to mark the film orien-tation Centre B used a thicker custom-built backscatter block (Fig 3 in Supplementary material) The different film exposure methods were compared
Film was analysed by the audit team using a bespoke ImageJ macro[18], which included measurements of field size and radia-tion field uniformity The 2 Gy film was used to measure the field size based on the 50% dose threshold defined by the 1 Gy film The reproducibility of the audit ImageJ macro field size analysis was within 1 pixel (0.25 mm) Radiation field uniformity was defined by the background-corrected ratio of the pixel value (PV)
at the four cardinal coordinates (above, below, left and right) to that at the centre of the field[18] Host film analysis methods var-ied from a uniformity calculation using the ratio (PVmax PVmin)/ (PVmax+ PVmin), to a more comprehensive bespoke Matlab routine that used a 2-channel (red and blue) analysis of the film and produced 2D horizontal and vertical profiles, a surface plot (3D profiles) and a 2D relative dose map
3 Results 3.1 Beam quality (HVL) All audit and host HVL measurements were within 10% of the baseline value established at commissioning by each centre Differ-ences in HVL jig design (seeSupplementary material) and measur-ing equipment resulted in audit/host discrepancies of up to 6.5%, 0.04 mmAl, in the measured HVL When assessing the effect of such HVL discrepancies in the determination of Dw only a 0.2% variation was observed (Fig 2)
3.2 Radiation output The main difference between the CoP recommendations and the host and audit output measurements was the use of the Ariane POM jig instead of measuring at the surface of a full scatter water-equivalent phantom Based on the comparison made by cen-tre A, the averages of the readings with the ionisation chamber in a full scatter block and in the output jig were within 0.4% of each other; this difference was not significant given the 0.8% relative standard deviation for the combined set of measurements and therefore no additional correction factor to the chamber factor,
kch, was introduced All absorbed dose to water measurements (both audit and host) were within ±2% of the expected value (30 Gy for a 3000 MU exposure) An agreement of better than 2.2% between auditor and host measurements was observed (Table 2) Strontium constancy checks before and after the audit agreed within 0.13%
Differences of up to 1°C or 0.4 °C were observed post-exposure between room temperature and chamber or jig temperatures, respectively This difference resulted in a 0.2% and 0.1% variation
in calculated radiation output, respectively No differences were observed pre-exposure between temperature measurement methods
3.3 Radiation field size and uniformity Host-audit field size measurements agreement was within
1 mm and both within 1 mm of the specified values Using the ImageJ macro, the median of the measured field uniformity in
Fig 1 Original output jig design; it is made of PMMA instead of POM and the
Trang 4terms of the relative difference between the field periphery and the
field centre mean pixel values was 2.7%
The back section of the output jig that included a 2 cm thick
backscatter block was the preference in all centres for film
mea-surements Field size measurements were up to 1% more accurate
when positioning the jig such that the backscatter block was flush
against the film Average uniformity across the field size was
improved by up to 1% with increased backscatter, i.e using centre
B’s custom-built backscatter block
3.4 Procedural audit
All centres had a well-established and documented Quality
Sys-tem in place and were generally in agreement with the QC testing
frequency and tolerances given by the IPEM 81 recommendations
for kV units and the practice followed by other electronic
brachytherapy users (Table 1) The clinical practice workload
var-ied across centres from less than 1 (centre B) to more than 4
(cen-tre A) patients per week Some correlation between confidence in
the machine performance and test frequency or tolerance levels
was observed In some cases tolerance levels were also linked to
the test methodology accuracy, for instance in film analysis
4 Discussion
All measurements, both by the host centres and the audit team,
were well within the IPEM 81 recommended tolerances for kV
units[13]and the tolerances used by other electronic
brachyther-apy users[6]
The UK CoP[14]recommends narrow beam geometry for HVL
measurements A kV regional audit of 70–300 kV in 2008, stated
an acceptable audit/host agreement limit of ±3.0%[8,9] However, the rapid dose fall-off of the Papillon 50 kV beam introduces more difficulties in the measurement of the HVL and, as shown by the results from this audit, a larger acceptable agreement limit should
be considered at such low energies A compromise is required between radiation scatter to the chamber, minimised with long applicator end to chamber window distances, and exposure times required for a reasonable signal-to-noise ratio Centre B used a custom-built POM applicator to increase the distance; this is in favour of a reduction of natural beam scatter but could be seen
as not representative of the clinical scenario Measurements with the chamber in PMMA provide better positioning stability but could introduce worse scatter conditions than in-air measure-ments Smaller beam collimator diameters in the lead plate, used
to reduce scatter, could on the other hand increase electron con-tamination to the chamber readings Dose rate and depth dose increase with applicator size due to increased scatter contribution
[10]; we recommend that an HVL jig be designed to allow investi-gation of the differences among all clinical field sizes Centre B excluded the machine ramp-up period from the HVL measure-ments by measuring charge for a fixed period of time while the
kV was stable This could have contributed to the host HVL being higher than that measured by the audit team Due to a machine overheating history at centre C, it was decided at the commission-ing that HVLs would be measured with exposures of 500 MU; the audit team followed this approach However, all other centres used exposures of 3000 MU, which potentially resulted in a higher signal-to-noise ratio thus reducing the uncertainty in the charge reading The beam profile is flat at the surface of the applicator end; however, due to the inherent rapid dose fall-off of the beam, beam hardness varies across the field size at a distance with the HVL at the periphery being lower than that at the centre If HVL
Fig 2 Graphical representation of the relative differences in HVL measurement between host and audit compared to the resulting differences in absorbed dose to water, D w , derived from variations in the selection of factors and coefficients for its calculation.
Table 2
Absorbed dose to water, D w , determined independently by host and audit for 3000 MU.
HVL (mmAl) Mass en abs coef ratio k ch D w (Gy) Audit vs host (% dif.)
Trang 5measurements are carried out following the UK CoP [14], the
resulting measured HVL could be higher than the mean HVL across
the treatment field size We recommend that a comparison be
made between the HVLs at the field centre and the periphery with
an assessment of the impact on the absorbed dose to water
determination
The 2008 kV regional audit protocol states an acceptable
host/audit radiation output agreement limit of 3.0%[8,9]; the same
limit could be applied for the Papillon 50 Different temperature
measurement methods did not introduce significant variations in
calculated radiation output However, we recommend that
tem-perature be measured as close to the chamber position as possible
A custom-built jig fitted in the chamber slot within the output jig
provides reproducibility of the measurement technique
Even though all centres produced radiation field size and
uni-formity results within recommended tolerances [13], the large
variation between film uniformity analysis methods did not allow
for a direct host-audit comparison All Papillon users would benefit
from an advanced film analysis providing 3D profile maps As the
energy independence of Gafchromic film has been doubted at
low energies[11,19], we recommend that film calibration be
per-formed at the same energy range as that of the Papillon radiation
beam for its use in dosimetric measurements
The dosimetric characterisation of the Papillon 50 was validated
with the audit results for all participating centres, thus providing
reassurance that the implementation had been performed within
the standards stated in previously published audit work and
recommendations for kV and electronic brachytherapy units
How-ever, it has also highlighted differences across the audited centres,
especially for the measurement of beam quality and the analysis of
field size and uniformity This audit should be considered as a
starting point for the development of national and international
guidelines for an optimised and standardised quality assurance
testing of the Papillon 50 unit
Acknowledgements
We would like to thank the staff from the following hospitals
for their cooperation and their kind hosting of this audit: Royal
Surrey County Hospital, Guildford; Clatterbridge Cancer Centre,
Liverpool; Nottingham University Hospital, Nottingham and Castle
Hill Hospital, Hull
We would also like to thank the National Physical Laboratory,
the NCRI Radiotherapy Trials Quality Assurance group and St
Tho-mas’ Hospital for providing the equipment used in this audit We
would like to acknowledge Matthew Bolt from Royal Surrey
County Hospital for producing the film analysis Image J macro
Finally, we would like to acknowledge Ariane Medical Systems
Ltd for funding and supporting this audit
Appendix A Supplementary data Supplementary data associated with this article can be found, in the online version, athttp://dx.doi.org/10.1016/j.phro.2016.12.001 References
[1] Clark CH, Aird EGA, Bolton S, et al Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.
Br J Radiol 2015;88:20150251 [2] Eaton DJ, Bolton S, Thomas RAS, Clark CH Interdepartmental dosimetry audits: development of methods and lessons learned J Med Phys 2015;40:183–9 [3] National Institute for Health and Care Excellence Low energy contact X-ray brachytherapy (the Papillon technique) for early stage rectal cancer: Interventional procedure guidance; 2015 http://www.nice.org.uk/guidance/ ipg532
[4] Palmer AL, Pearson M, Whittard P, McHugh KE, Eaton DJ Current status of kilovoltage radiotherapy in the UK: installed equipment, clinical workload, physics quality control and radiation dosimetry Br J Radiol 2016 http://dx.doi org/10.1259/bjr.20160641
[5] Myint AS Radiotherapy for early rectal cancer Clin Oncol 2007;19:637–8 [6] Eaton DJ Electronic brachytherapy – current status and future directions Br J Radiol 2015;88:20150002
[7] Budgell G, Berresford J, Trainer M, Bradshaw E, Sharpe P, Williams P A national dosimetric audit of IMRT Radiother Oncol 2011;99:246–52
[8] Palmer A, Mzenda B, Kearton J, Wills R Analysis of regional radiotherapy dosimetry audit data and recommendations for future audits Br J Radiol 2011;84:733–42
[9] Burton NLA, Brimelow J, Welsh AD A regional audit of kilovoltage X-rays – a single centre approach Br J Radiol 2008;81:422–6
[10] Gérard J-P, Myint AS, Groce O, et al Renaissance of contact x-ray therapy for treating rectal cancer Expert Rev Med Devices 2011;8(4):483–92
[11] Croce O, Hachem S, Franchisseur E, Marcié S, Gérard J-P, Bordy J-M Contact radiotherapy using a 50 kV x-ray system: evaluation of relative dose distribution with the Monte Carlo code PENELOPE and comparison with measurements Radiat Phys Chem 2012;81:609–17
[12] Blake SW, Casebow MP A pragmatic approach to dosimetric audit in radiotherapy Br J Radiol 2002;75:754–62
[13] Mayles WPM, Lake R, McKenzie A, et al IPEM Report 81: Physics Aspects of Quality Control in Radiotherapy The Institute of Physics and Engineering in Medicine; 1999
[14] Klevenhagen SC, Aukett RJ, Harrison RM, Moretti C, Nahum A, Rosser KE The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al – 4 mm Cu HVL: 10–300 kV generating potential) Phys Med Biol 1996;41:2605–25
[15] Aukett RJ, Burns JE, Greener AG, et al Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al – 4 mm Cu HVL) Phys Med Biol 2005;50:2739–48 [16] Ipe NE, Rosser KE, Moretti CJ, Manning JW, Palmer MJ Air kerma calibration factors and chamber correction values for PTW soft x-ray, NACP and Roos ionization chambers at very low x-ray energies Phys Med Biol 2001;46: 2107–17
[17] Perrin BA, Whitehurst P, Cooper P, Hounsell AR The measurement of k ch factors for application with the IPEMB very low energy dosimetry protocol Phys Med Biol 2001;46:1985–95
[18] Bolt M Radiation field size and uniformity analysis ImageJ macro instructions (Royal Surrey County Hospital).
[19] Allen Li X, Ma C-M, Salhani D, Agboola O Dosimetric evaluation of a widely used kilovoltage x-ray unit for endocavitary radiotherapy Am Assoc Phys Med
1998 S0094-2405(98)01808-2