CONTRAST MEDIAFeasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study Robbert W.. This article is published with op
Trang 1CONTRAST MEDIA
Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
Robbert W van Hamersvelt1&Martin J Willemink1&Pim A de Jong1&Julien Milles2&
Alain Vlassenbroek3&Arnold M R Schilham1&Tim Leiner1
Received: 11 August 2016 / Revised: 12 December 2016 / Accepted: 3 January 2017
# The Author(s) 2017 This article is published with open access at Springerlink.com
Abstract
Objectives The aim of this study was to evaluate the
feasibil-ity and accuracy of dual-layer spectral detector CT (SDCT) for
the quantification of clinically encountered gadolinium
concentrations
Methods The cardiac chamber of an anthropomorphic
tho-racic phantom was equipped with 14 tubular inserts
con-taining different gadolinium concentrations, ranging from
0 to 26.3 mg/mL (0.0, 0.1, 0.2, 0.4, 0.5, 1.0, 2.0, 3.0, 4.0,
5.1, 10.6, 15.7, 20.7 and 26.3 mg/mL) Images were
ac-quired using a novel 64-detector row SDCT system at 120
and 140 kVp Acquisitions were repeated five times to
assess reproducibility Regions of interest (ROIs) were
drawn on three slices per insert A spectral plot was
ex-tracted for every ROI and mean attenuation profiles were
fitted to known attenuation profiles of water and pure
gadolinium using in-house-developed software to
calcu-late gadolinium concentrations
Results At both 120 and 140 kVp, excellent correlations
between scan repetitions and true and measured
gadolin-ium concentrations were found (R > 0.99, P < 0.001;
ICCs > 0.99, CI 0.99–1.00) Relative mean measurement
errors stayed below 10% down to 2.0 mg/mL true
gadolinium concentration at 120 kVp and below 5% down
to 1.0 mg/mL true gadolinium concentration at 140 kVp Conclusion SDCT allows for accurate quantification of gad-olinium at both 120 and 140 kVp Lowest measurement errors were found for 140 kVp acquisitions
Key Points
• Gadolinium quantification may be useful in patients with contraindication to iodine
• Dual-layer spectral detector CT allows for overall accurate quantification of gadolinium
• Interscan variability of gadolinium quantification using SDCT material decomposition is excellent
Keywords Dual-energy CT Dual-layer spectral detector CT Contrast media Gadolinium Material decomposition
Abbreviations
ICC Intraclass correlation coefficient
NIST National Institute of Standards and Technology
SDCT Dual-layer spectral detector computed tomography
* Robbert W van Hamersvelt
R.W.vanHamersvelt-3@umcutrecht.nl
1
Department of Radiology, University Medical Center Utrecht,
P.O Box 85500, 3508 GA Utrecht, The Netherlands
2 CT Clinical Science, Philips HealthCare, Best, The Netherlands
3
CT Clinical Science, Philips HealthCare, Brussels, Belgium
DOI 10.1007/s00330-017-4737-8
Trang 2Material decomposition imaging (MDI) using dual-energy
computed tomography (DECT) was first described by
Hounsfield in 1973 [1] Different materials, which cannot be
distinguished on the basis of attenuation number, can be
dis-tinguished with the use of material decomposition algorithms
using DECT acquisitions [2–4] Materials with high atomic
numbers, such as iodine (Z = 53) and gadolinium (Z = 64),
show characteristic high attenuation profiles at low energies
owing to a substantial contribution of the photoelectric effect
to the attenuation [5] MDI uses these characteristic
attenua-tion profiles to differentiate these contrast agents from other
materials MDI has not been widely applied in clinical practice
until recently Over the past few years several CT vendors
have made DECT commercially available for daily clinical
practice Recently a novel DECT technique has become
com-mercially available, which uses a single tube with a dual-layer
detector capable of differentiating between low and high
en-ergy X-ray photons, and is further investigated in this study
One of the most widely researched MDI applications is
quan-titative mapping of iodine distribution in tissues The resulting
maps can be used as a surrogate for tissue perfusion Early
evi-dence has shown the clinical capability of iodine quantification
with DECT at a specified time point for the detection of
myocar-dial [6–12] and pulmonary perfusion defects [13–16] In
addi-tion, DECT iodine mapping is capable of tumour mass
charac-terization and therapy response assessment [17–19] However,
iodine contrast administration, while safe in most patients, is
associated with contrast-induced allergic reactions and
nephrop-athy which can cause acute renal dysfunction [20,21] and
sig-nificant morbidity and mortality, especially in high-risk patients
[22,23] In patients with contraindications to iodinated contrast
media, gadolinium-enhanced magnetic resonance (MR)
angiog-raphy can be used as an alternative However, depending on the
indication, MR angiography may have poor diagnostic value
compared to (DE)CT angiography Gadolinium-based CT has
been used off-label in higher doses as an alternative for
conven-tional CT angiography with diagnostic image quality [24,25]
With the use of DECT, higher attenuation can be achieved at low (monochromatic) energies, which could enable the use of much lower gadolinium concentrations [26,27] In addition, accurate gadolinium quantification using DECT could allow for a quanti-tative evaluation of contrast agent distribution in tissue as a sur-rogate for tissue perfusion using MDI Therefore, accurate gad-olinium quantification combined with increased attenuation could potentially open up the possibility for gadolinium as an alternative contrast agent for DECT imaging in patients with contraindications to iodinated contrast media
In several studies the feasibility of gadolinium-enhanced DECT has been reported in phantom and animal models [28–31] These studies described the capability of spectral dif-ferentiation and visualisation [28–30] and accuracy of quantifi-cation [31] of gadolinium using DECT However, the accuracy
of gadolinium quantification using the novel dual-layer spectral detector CT system (SDCT) is unknown Therefore, the aim of the current study was to evaluate the feasibility and accuracy of gadolinium quantification using a SDCT system
Materials and methods Phantom design
An anthropomorphic chest phantom (QRM GmbH, Moehrendorf, Germany) was used The phantom resembles a chest with corresponding X-ray attenuation behaviour The phan-tom has a cylindrical cardiac chamber in which a plastic holder was placed (Fig.1) Three plastic holders were made, two consisting of five tubular inserts, and one consisting of three tubular inserts with surrounding 2% agar gel solution In addi-tion, a plastic holder with one tubular insert containing water with surrounding 2% agar gel solution served as control The fourteen 32-mL tubular inserts contained different concentrations of the gadolinium-based contrast agent gadobutrol (Gadovist 1.0, Bayer Healthcare, Berlin, Germany) One millilitre of this con-trast agent contains 157.25 mg gadolinium Different amounts of gadobutrol were diluted in water, resulting in the following
Fig 1 Phantom setup a
Anthropomorphic thoracic
phantom with a plastic holder
placed in the cardiac chamber b
Representative plastic holder
filled with 5 tubular inserts, with
surrounding 2% agar gel solution
Trang 3concentrations of gadolinium: 0.0, 0.1, 0.2, 0.4, 0.5, 1.0, 2.0, 3.0,
4.0, 5.1, 10.6, 15.7, 20.7 and 26.3 mg/mL, which is equivalent to
0.000, 0.001, 0.002, 0.002, 0.003, 0.006, 0.013, 0.019, 0.026,
0.032, 0.068, 0.100, 0.132 and 0.167 mmol/mL, respectively
Concentrations were chosen to mimic an estimated clinical
range of gadolinium concentrations encountered after
injec-tion of 0.1–0.2 mmol of gadolinium per kilogram Strich et al
[32] measured percentage dose of gadolinium-based contrast
agent per gram of tissue in healthy rabbit organs 5 min after
admission, resulting in the following percentages: 0.052%/g
heart, 0.073%/g lungs, 0.037%/g liver, 0.037%/g spleen and
0.250%/g kidney On the basis of these percentages, an
esti-mation of gadolinium concentrations encountered at each
or-gan can be calculated At 31.5 mg/kg bodyweight (equal to
0.2 mmol/kg) gadolinium administration, a human subject of
70 kg would receive a total of 2201.5 mg gadolinium On the
basis of the percentages determined by Strich and colleagues,
gadolinium distribution in the heart 5 min after injection
would be 0.00052 × 2201.5 mg, or 1.15 mg per gram
myocar-dium Myocardial muscle has a specific gravity of 1.05 g/mL
[33], implicating an estimated gadolinium concentration
en-countered in the myocardium of 1.15 mg/g × 1.05 g/mL, or
1.21 mg/mL Using the aforementioned distribution
percent-ages these calculations can also be applied to the lungs, liver,
spleen and kidney, with a calculated estimated specific gravity
(weight/volume) of 1.34, 1.01, 0.71 and 0.85 g/mL,
respec-tively [34–36] Thus, it is to be expected that gadolinium
concentrations of 2.15, 0.82, 0.58 and 4.67 mg/mL are
en-countered in healthy lung, liver, spleen and kidney tissue,
respectively These concentrations are in the range of
concen-trations evaluated in this study As it is expected that in tissue
with a perfusion defect lower concentrations of gadolinium
will be encountered, we also evaluated ultra-low
concentra-tions of gadolinium down to 0.1 mg/mL
Image acquisition
Images were acquired using the newest generation 64-detector
row SDCT system (iQon Spectral CT, Philips Healthcare,
Best, the Netherlands) This system uses a single X-ray tube
and a dual-layer detector The detector separates the X-ray
beam into low (upper layer) and high (lower layer) energy
data, which is used to reconstruct spectral-based images
(SBI) The SBI contain the raw data of both layers and are
used to reconstruct any dual-energy image and/or analysis In
addition, by combining the output of both layers, a
conven-tional image is reconstructed from the data The phantom was
imaged in spiral mode at 120 and 140 kVp The tube current–
time product was set to a fixed value of 200 mAs, resulting in
a volumetric CT dose index (CTDIvol) of 18.4 and 26.5 mGy
for 120 and 140 kVp acquisitions, respectively The following
parameters were used: detector collimation 64 × 0.625 mm,
rotation time 0.4 s and pitch 1.046 At both tube voltages,
acquisitions were repeated five times with small displace-ments between each acquisition to take into account interscan variation Thus, the phantom was translated a few millimetres
in the left–right direction, as well as along the z-axis of the CT scanner After the five repetitions, the phantom was set back to the starting position
Image reconstruction The raw projection data from both detector layers were auto-matically reconstructed into SBI Subsequently, MDI was per-formed in the projection domain, which efficiently eliminates beam hardening artefacts [37] All images were reconstructed with standard chest reconstruction filter B and spectral level 3 Spectral is a model-based iterative reconstruction developed for the SDCT, it is an equivalent to iterative model-based reconstruction (IMR) Spectral consists of six levels, whereby
a higher spectral level implies more noise reduction Slice thickness and increment were both 1 mm The reconstructed images were evaluated on a dedicated workstation using the Spectral CT Viewer (IntelliSpace Portal v6.5.0.02080, Philips Healthcare, Best, the Netherlands)
Image analysis and gadolinium quantification
On three different slices of each data set a region of interest (ROI) with a fixed size of 225 mm2was drawn in the centre of each insert (Fig.2a) Subsequently spectral plots of every ROI were obtained, in which mean Hounsfield units (HU) were plotted as a function of different energy levels expressed in kilo electron volt (keV) (Fig.2b) These mean HU values of the spectral plots were extracted in steps of 10 keVand used as
an input for the analysis The currently used SDCT system uses traditional integrated detectors at two energy spectra and is therefore not able to image and/or quantify a material-specific K-edge [37] Materials with a K-edge within the SDCT range (40–200 keV) will not show a discontinuity in their attenuation function on the SDCT spectral plot When evaluating the mean attenuation across monochromatic ener-gies, this does not pose a problem and therefore the whole energy spectrum can be used (40–200 keV) However, for the quantitative analyses of gadolinium concentrations a com-parison is made with the attenuation profile of pure
gadolini-um which does contain the discontinuity in their attenuation function at the K-edge Therefore, to take into account the non-linear energy dependency close to the K-edge of gadolin-ium (50.2 keV), only the energy range from 70 to 200 keV was used for the quantitative analysis With in-house-developed software, attenuation profiles were reconstructed from the provided mean HU, and gadolinium concentrations were calculated by fitting combinations of known attenuation profiles of pure gadolinium and water to the
reconstruct-ed attenuation profiles (Fig.2c) For each ROI drawn in the
Trang 4phantom, the in-house-developed software assumed that all
voxels within this ROI were composed of only gadolinium
and water and that the sum of these fractions added up to
100% Known attenuation profiles of pure gadolinium and
water were obtained from the National Institute of
Standards and Technology (NIST) database [38] Therefore,
no calibration scans with water and/or gadolinium
concentra-tions were needed For all thirteen different gadolinium
con-centrations, 15 measurements were performed at both 120 and
140 kVp (three slices, five repetitions) In addition, 30
mea-surements (15 at both 120 and 140 kVp) were performed on
the control phantom Gadolinium concentrations were
calcu-lated for each measurement
Attenuation coefficient
CT attenuation during injection of low gadolinium concentra-tions (i.e 0.1–0.2 mmol/kg bodyweight) will generally lead to lower HU values compared to the use of iodinated contrast agents [24,25] To investigate the ability of SDCT to visually identify an increase in HU values due to the presence of a gadolinium-containing contrast agent we extracted mean at-tenuation coefficients across monochromatic energies (40–
200 keV) for the different gadolinium concentrations used in this study (Fig.3)
Statistical analysis
To evaluate the quantification accuracy of gadolinium concen-trations, we defined measurement errors in milligrams per millilitre and relative measurement errors in percentages Measurement errors were calculated by subtracting true gad-olinium concentrations from the measured gadgad-olinium con-centrations Subsequently, relative measurement errors (%) were calculated as follows:
Relative measurement errorð Þ%
mg mL
true gadolinium concentration mg
mL
100 %ð Þ
All measurement error analyses were performed sepa-rately for 120 and 140 kVp In addition, sub-analyses were done for each concentration The Shapiro–Wilk test was used to identify normally distributed data For each con-centration, statistical differences of measurement errors between 120 and 140 kVp were analysed using paired t test for normally distributed data A Bonferroni corrected
P < 0.004 (0.05/number of comparisons) was considered significant Pearson’s correlation coefficient was used to evaluate correlations between measured and true gadolin-ium concentrations at different tube voltages and for each scan repetition In addition, reproducibility was evaluated
To define agreement of results, the two-way random single measure intraclass correlation coefficient (ICC) with cor-responding confidence interval (CI) was used for all pos-sible two-way interactions ICCs between 0.61 and 0.80 were considered good and ICCs greater than 0.80 excel-lent [39] Measurement interscan variabilities of all scan repetitions were plotted in one single plot by using a mod-ified Bland–Altman plot described by Jones et al [40] In this figure the measurement differences of every measure-ment compared to the mean measuremeasure-ment of all scans are plotted against the mean measurement of all scans As described by Jones et al., the limits of agreement were calculated as mean ± 1.96 × SD, where the SD is an
Fig 2 Axial CT image and measurements a Axial conventional SDCT
image of the phantom with 5 tubular inserts, surrounded by 2% agar gel.
ROIs with a fixed area of 225 mm 2 drawn in the centre of each insert b A
spectral plot of each ROI was conducted, showing mean Hounsfield units
plotted against energy in keV Hounsfield unit values of the spectral plots
were extracted in increments of 10 keV c Using in-house-developed
software, we reconstructed attenuation profiles between 70 to 200 keV
from the extracted Hounsfield units, and a combination of known
attenuation profiles of pure gadolinium and water was fitted to the
reconstructed attenuation profile This case concerns ROI S3, containing
5.1 mg of gadolinium per millilitre
Trang 5estimate of the standard deviation for all observers [40].
Values are listed as mean with standard deviation (SD),
unless stated otherwise A P value less than 0.05 was used
to indicate statistical significance IBM SPSS version 21.0
(IBM corp., Armonk, New York, USA) was used for
sta-tistical analyses
Results
Measurements of the water-filled insert, which served as
control, yielded 0.0 ± 0.0 mg/mL with a measurement
error of 0.0 ± 0.0 mg/mL for all measurements To avoid
influence on measurement accuracy, these control
measurements were not included in further statistical analyses
Accuracy and reproducibility
At both 120 and 140 kVp, excellent correlations (R > 0.99,
P < 0.001; ICCs > 0.99, CI 0.99–1.00) were found between true and measured gadolinium concentrations for each scan repetition In addition, reproducibility between all scan repe-titions was excellent (R > 0.99, P < 0.001; ICCs > 0.99, CI 0.99–1.00) The interscan agreement is displayed in Fig.4a for 120 kVp and Fig 4b for 140 kVp Because excellent correlations were found, all scan repetitions were analysed combined together in subsequent analyses
0 250 500 750 1000 1250 1500 1750 2000 2250 2500
40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200
40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200
Energy (keV)
26.3 mg/mL 20.7 mg/mL 15.7 mg/mL 10.6 mg/mL 5.1 mg/mL 4.0 mg/mL 3.0 mg/mL 2.0 mg/mL 1.0 mg/mL 0.5 mg/mL 0.4 mg/mL 0.2 mg/mL 0.1 mg/mL
Gadolinium concentration
Mean attenuation 120 kVp 2750
0 250 500 750 1000 1250 1500 1750 2000 2250 2500 2750
Energy (keV)
26.3 mg/mL 20.7 mg/mL 15.7 mg/mL 10.6 mg/mL 5.1 mg/mL 4.0 mg/mL 3.0 mg/mL 2.0 mg/mL 1.0 mg/mL 0.5 mg/mL 0.4 mg/mL 0.2 mg/mL 0.1 mg/mL
Gadolinium concentration
Mean attenuation 140 kVp
a
b
Fig 3 Mean CT attenuation
coefficients across all
monochromatic energies Mean
CT attenuation of all
measurements for each
gadolinium concentration,
constructed in steps of 10 keV.
Graphs were used to investigate
the ability of SDCT low
monochromatic energies to
visually identify an increase in
HU values due to the presence of
gadolinium-containing contrast
media Scans were performed at
120 kVp (a) and 140 kVp (b) For
subsequent gadolinium
quantification, only attenuation
profiles between 70 to 200 keV
were used for the
in-house-developed software analyses
(Fig 2c )
Trang 6120 kVp
All gadolinium concentrations were overestimated Mean
measurement errors for the 15 ROIs per concentration ranged
between 0.1 and 2.4 mg/mL (Table1, Fig.5a) For each
con-centration, measurement errors at 120 kVp were significantly
(Bonferroni P < 0.004) higher compared to measurement
er-rors at 140 kVp, except for the lowest two concentrations of
0.1 and 0.2 mg/mL Relative measurement errors (%) were
below 10% down to 2.0 mg/mL true gadolinium
concentra-tions and increased up to 29.4% at 0.5 mg/mL and 100.9% at
0.1 mg/mL true gadolinium concentration (Table1, Fig.5b)
140 kVp
Per concentration (N = 15), mean measurement errors varied
from−0.2 to 0.4 mg/mL (Table1, Fig.5a) Relative
measure-ment errors (%) stayed below 5% down to 1.0 mg/mL true
gadolinium concentration At true gadolinium concentrations
between 0.1 and 0.5 mg/mL, mean measurement errors were low with 0.1 ± 0.0 mg/mL deviation; expressed in percentages this varied between 93.5 ± 26.8% and 14.1 ± 4.1% deviation, respectively (Fig.5b)
Attenuation coefficient Overall mean attenuation increased when lowering keV (Fig 3) At the lowest possible monochromatic energy (40 keV), mean attenuation for the estimated clinical gadolin-ium range of 0.5, 1.0, 2.0, 3.0, 4.0 and 5.1 mg/mL yielded 28,
74, 164, 260, 349 and 416 HU at 120 kVp and 34, 84, 180,
284, 382 and 464 HU at 140 kVp, respectively
Discussion This study showed that it is feasible to quantify a commonly clinically encountered range of gadolinium concentrations in a
-0.2 -0.15 -0.1 -0.05 0 0.05 0.1
0.15
Interscan variation 120 kVp
+1.96 SD
-1.96 SD
0.08
-0.08 Scan 1
Scan 2 Scan 3 Scan 4 Scan 5
Mean gadolinium concentration measurement of all observers
-0.5 -0.4 -0.3 -0.2 -0.1 0 0.1 0.2 0.3
Mean gadolinium concentration measurement of all observers
Interscan variation 140 kVp
Scan 1 Scan 2 Scan 3 Scan 4 Scan 5
-1.96 SD
+1.96 SD
-0.14 0.14
a
b
Fig 4 Interscan agreement for all
scan repetitions at 120 kVp (a)
and 140 kVp (b) Values are
plotted according to Jones et al.
[ 40 ] The measurement difference
of each scan compared to the
mean measurement of all scans is
plotted against the mean
measurement of all scans
Trang 7phantom model with overall high accuracy and reproducibility
using an in-house-developed material decomposition method
on a novel clinical dual-layer spectral detector CT system
Whereas conventional CT displays anatomical structures
as a function of tissue density, DECT enables enhanced tissue
characterization using MDI Quantitative assessment of
con-trast agent uptake and its provided distribution map can be
used as a surrogate for tissue perfusion [6–12,14] In the
current study we showed that clinically encountered low
con-centrations of gadolinium, down to 0.5 mg/mL, can be
accu-rately quantified with a mean measurement error of 0.1 mg/
mL using SDCT at both 120 and 140 kVp In the ultra-low
gadolinium concentration range (0.1–0.4 mg/mL), expected to
be encountered in tissues with a perfusion defect, the mean
measurement error remained around 0.1 mg/mL at both 120
and 140 kVp However, at these low concentrations the
mar-gin of error increased substantially and approached the
gado-linium concentration itself, indicating that the lower limit of
reasonably accurate gadolinium quantification using SDCT
lies between 0.5 and 1.0 mg/mL In the range of clinically
encountered gadolinium concentrations (0.5–5.1 mg/mL)
af-ter administration of 0.1–0.2 mmol/kg bodyweight, mean CT
numbers at 40 keV ranged between 28 and 464 HU (Fig.3)
The combination of high(er) attenuation at lower
monochro-matic energies and accurate quantification of low gadolinium
concentrations opens up the possibilities for DECT scanning
with the use of gadolinium as a contrast agent Potential
clin-ical applications include detection of myocardial [6–12] and
pulmonary perfusion defects [14–16] and the characterization
of tumour masses and therapy response assessment [17–19]
In clinical routine, adequate tissue contrast and contrast agent density maps are important for the diagnosis and evalu-ation of organ perfusion defects However, to be able to create
a gadolinium density map as a surrogate for tissue perfusion, accurate gadolinium quantification is essential, as the post-processing is based on these measurements This is the first study to describe the accuracy of gadolinium quantification using MDI on SDCT Gabbai et al [28] described the capabil-ity of spectral differentiation of gadolinium using SDCT, which is in accordance with our study However, no quantita-tive values were described and high concentrations (4.7– 187.6 mg/mL) of gadolinium were used, which is at least one
to two orders of magnitude above the estimated range encoun-tered in healthy cardiac, lung, liver, spleen and kidney tissue (0.58–4.66 mg/mL) Zhang et al [30] showed a high sensitiv-ity and specificsensitiv-ity for gadolinium-enhanced dual-source DECT pulmonary angiography to detect pulmonary embolism
in rabbits However, as in the study by Gabbai et al gadolin-ium concentration was not quantified In addition, high intra-venous doses of gadolinium contrast agent, 1.5 and 2.5 mmol/
kg bodyweight, were administrated Bongers et al [31] evalu-ated the potential of gadolinium as a CT contrast agent using dual-source DECT in a phantom setup In accordance with our study they found that monochromatic images at low energy (e.g 40 keV) allow for higher attenuation Additional quanti-fication was performed by using the material-specific
dual-Table 1 Mean errors of
gadolinium concentration
measurements with a dual-layer
spectral detector CT scanner
True concentration (mg/mL) 120 kVp 140 kVp
Measurement error Measurement error
Data are given as mean ± standard deviation For each true concentration 15 measurements were done at both 120 and 140 kVp
*Significantly (Bonferroni P < 0.004) higher compared to measurement error at 140 kVp
Trang 8energy ratio for gadolinium For the true gadolinium
concen-trations 6.3, 3.2, 1.6, 0.8, 0.4 and 0.2 mg/mL relative
measure-ment errors were 11.5, 12.0, 21.6, 21.6, 104.2 and 159.4%,
respectively In our study we found a higher accuracy with
relative measurement errors of less than 10% down to
2.0 mg/mL at 120 kVp and 1.0 mg/mL at 140 kVp A possible
explanation for this difference can be found in the algorithm
The post-processing algorithms used by Bongers et al [31] was
originally designed for iodine, whereas our algorithm was
spe-cifically designed for gadolinium quantification
We found a slightly lower measurement error, and thus
higher accuracy, for scans acquired at 140 kVp compared to
120 kVp When scanning with a higher tube voltage, more
high energy X-ray photons are produced This decreases the
spectral overlap between high- and low-energy spectra, and
thereby improves the accuracy of material decomposition,
which is in accordance with the findings of Gabbai and
col-leagues [28] Moreover, 140 kVp acquisitions resulted in
higher CT numbers of different gadolinium concentrations at
monochromatic 40 keV images (34–464 HU) compared to
120 kVp acquisitions (28 to 416 HU), indicating a superior spectral separation at a higher tube voltage
Even though gadolinium chelates are generally considered to
be safe contrast agents, with acute reaction rates of approximately 0.001–0.07% [41], recently concerns have arisen about their long-term safety after the discovery that administration of multi-ple doses has led to detectable gadolinium levels in the brain [42,
43] In addition, gadolinium contrast has been linked to an in-creased risk of nephrogenic systemic fibrosis (NSF) in patients with impaired renal function [44] In both conditions the linear non-ionic and linear ionic contrast agents have primarily been implicated, whereas macrocyclic gadolinium agents, such as used in the current study, have not been linked conclusively to either of these conditions [45–47] Although both iodine and gadolinium contrast agents pose a risk for patients with impaired renal function, gadolinium is thought to be preferred in patients with renal failure and a glomerular filtration rate greater than
30 mL/min since the risk of NSF is low in these patients, while the risk of iodine contrast-induced nephropathy clearly exists [41] Furthermore, using gadolinium could potentially obviate the need for pre- and post-imaging hydration as well as premedication protocols that are commonly used in patients with impaired renal function who undergo contrast-enhanced CT scanning, or patients with known allergies to iodinated contrast agents In the current study a relatively simple method for mate-rial decomposition using in-house-developed software is pro-posed Our method is based on the mass attenuation coefficient across monochromatic energies Monochromatic reconstructions take into account the function of two independent factors: the photoelectric and the Compton effect [2] The photoelectric effect
is strongly related to the atomic number of a material in the CT energy range and is therefore material-specific [37] Our method takes into account this material-specific effect by evaluating the attenuation across monochromatic energies
The strength of our study is that we evaluated accuracy of gadolinium quantification in an optimal controlled setting with a wide and clinically relevant range of gadolinium con-centrations, which provides the basis for further research and clinical applications Our study also has some limitations The most important is that we used a static phantom in which organ motion was not taken into account In addition, a fixed concentration is not the same as a bolus injection However,
we tried to mimic the clinical situation as best as possible by using low concentrations of gadolinium, which are expected
to be typically encountered clinically A second limitation is that our study only takes into account water and gadolinium when calculating the amount of gadolinium concentration Since human tissue does not only consist of water and gado-linium, future phantom and patient research will have to ad-dress (healthy) tissue attenuation as well using a three- or multi-material decomposition method A third limitation is the need for relatively high peak tube voltage (120 or
a
b
Fig 5 Accuracy of gadolinium quantification Accuracy expressed as
mean measurement error (a) and mean relative measurement error (b).
Symbol represents mean and error bar the standard deviation
Trang 9140 kVp) settings to ensure sufficient spectral separation.
However, the higher radiation dose due to the use of
high kVp acquisitions can be addressed by reducing tube
cur-rent (mAs) The fourth limitation is that we only evaluated one
DECT technique; therefore, our results may be limited to the
vendor used in this study
In conclusion, SDCT allows for accurate quantification of
commonly clinically used gadolinium concentrations at both
120 and 140 kVp Lowest measurement errors were found for
140 kVp acquisitions
Compliance with ethical standards
Guarantor The scientific guarantor of this publication is Prof T.
Leiner.
Conflict of interest The authors of this manuscript declare
relation-ships with the following companies: Alain Vlassenbroek and Julien
Milles are employees of Philips Healthcare.
The other authors of this manuscript declare no relationships with any
companies whose products or services may be related to the subject
mat-ter of the article.
Funding The authors state that this work has not received any funding.
Statistics and biometry One of the authors has significant statistical
expertise.
Informed consent Written informed consent was not required because
this concerns a phantom study.
Ethical approval Institutional review board approval was not required
because this concerns a phantom study.
Methodology
• prospective
• experimental
• performed at one institution
Open Access This article is distributed under the terms of the Creative
C o m m o n s A t t r i b u t i o n 4 0 I n t e r n a t i o n a l L i c e n s e ( h t t p : / /
creativecommons.org/licenses/by/4.0/), which permits unrestricted use,
distribution, and reproduction in any medium, provided you give
appro-priate credit to the original author(s) and the source, provide a link to the
Creative Commons license, and indicate if changes were made.
References
1 Hounsfield GN (1973) Computerized transverse axial scanning
(to-mography) 1 Description of system Br J Radiol 46:1016 –1022
2 Alvarez RE, Macovski A (1976) Energy-selective reconstructions
in X-ray computerized tomography Phys Med Biol 21:733 –744
3 Macovski A, Alvarez RE, Chan JL et al (1976) Energy dependent
reconstruction in X-ray computerized tomography Comput Biol
Med 6:325–336
4 den Harder AM, Willemink MJ, de Jong PA et al (2016) New
horizons in cardiac CT Clin Radiol 71:758 –767
5 Johnson TRC, Krauss B, Sedlmair M et al (2007) Material differ-entiation by dual energy CT: initial experience Eur Radiol 17:
1510 –1517
6 So A, Hsieh J, Narayanan S et al (2012) Dual-energy CT and its potential use for quantitativemyocardial CT perfusion J Cardiovasc Comput Tomogr 6:308 –317
7 Ko SM, Choi JW, Song MG et al (2011) Myocardial perfusion imaging using adenosine-induced stress dual-energy computed to-mography of the heart: comparison with cardiac magnetic reso-nance imaging and conventional coronary angiography Eur Radiol 21:26 –35
8 Ruzsics B, Lee H, Zwerner PL et al (2008) Dual-energy CT of the heart for diagnosing coronary artery stenosis and myocardial ischemia-initial experience Eur Radiol 18:2414 –2424
9 Schwarz F, Ruzsics B, Schoepf UJ et al (2008) Dual-energy CT of the heart —principles and protocols Eur J Radiol 68:423–433
10 Fornaro J, Leschka S, Hibbeln D et al (2011) Dual- and multi-energy CT: approach to functional imaging Insights Imag 2:149 –159
11 Danad I, Fayad ZA, Willemink MJ, Min JK (2015) New applica-tions of cardiac computed tomography JACC Cardiovasc Imaging 8:710 –723
12 Zhang L-J, Peng J, Wu S-Y et al (2010) Dual source dual-energy computed tomography of acute myocardial infarction: correlation with histopathologic findings in a canine model Invest Radiol 45:
290 –297
13 Zhang L-J, Chai X, Wu S-Y et al (2009) Detection of pulmonary embolism by dual energy CT: correlation with perfusion scintigraphy and histopathological findings in rabbits Eur Radiol 19:2844 –2854
14 Bauer RW, Frellesen C, Renker M et al (2011) Dual energy CT pulmonary blood volume assessment in acute pulmonary embolism
- correlation with D-dimer level, right heart strain and clinical out-come Eur Radiol 21:1914 –1921
15 Krissak R, Henzler T, Reichert M et al (2010) Enhanced visualiza-tion of lung vessels for diagnosis of pulmonary embolism using dual energy CT angiography Invest Radiol 45:341 –346
16 Wu HW, Cheng JJ, Li JY et al (2012) Pulmonary embolism detec-tion and characterizadetec-tion through quantitative iodine-based material decomposition images with spectral computed tomography imag-ing Invest Radiol 47:85 –91
17 Baxa J, Vondráková A, Matou šková Tet al (2014) Dual-phase dual-energy CT in patients with lung cancer: assessment of the additional value of iodine quantification in lymph node therapy response Eur Radiol 24:1981 –1988
18 Knobloch G, Jost G, Huppertz A et al (2014) Dual-energy
comput-ed tomography for the assessment of early treatment effects of re-gorafenib in a preclinical tumor model: comparison with dynamic contrast-enhanced CT and conventional contrast-enhanced single-energy CT Eur Radiol 24:1896 –1905
19 Apfaltrer P, Meyer M, Meier C et al (2012) Contrast-enhanced dual-energy CT of gastrointestinal stromal tumors: is iodine-related attenua-tion a potential indicator of tumor response? Invest Radiol 47:65 –70
20 Lameire N, Van Biesen W, Vanholder R (2005) Acute renal failure Lancet 365:417 –430
21 Singri N, Ahya SN, Levin ML (2003) Acute renal failure JAMA 289:747 –751
22 Tepel M, Aspelin P, Lameire N (2006) Contrast-induced nephrop-athy: a clinical and evidence-based approach Circulation 113:
1799 –1806
23 Hunt CH, Hartman RP, Hesley GK (2009) Frequency and severity
of adverse effects of iodinated and gadolinium contrast materials: retrospective review of 456,930 doses AJR Am J Roentgenol 193:
1124 –1127
24 Esteban JM, Alonso A, Cervera V, Martínez V (2007) One-molar gadolinium chelate (gadobutrol) as a contrast agent for CT angiog-raphy of the thoracic and abdominal aorta Eur Radiol 17:2394 – 2400
Trang 1025 Remy-Jardin M, Dequiedt P, Ertzbischoff O et al (2005) Safety and
effectiveness of gadolinium-enhanced multi-detector row spiral CT
angiography of the chest: preliminary results in 37 patients with
contraindications to iodinated contrast agents Radiology 235:
819 –826
26 FitzGerald PF, Colborn RE, Edic PM et al (2016) CT image contrast
of high-Z elements: phantom imaging studies and clinical
implica-tions Radiology 278:723 –733
27 Thomas JV, Bolus DN, Jackson BE et al (2016) Gadoxetate
disodium enhanced spectral dual-energy CT for evaluation of
chol-angiocarcinoma: preliminary data Ann Med Surg 6:17 –22
28 Gabbai M, Leichter I, Mahgerefteh S, Sosna J (2015) Spectral
ma-terial characterization with dual-energy CT: comparison of
com-mercial and investigative technologies in phantoms Acta Radiol
56:960 –969
29 Fält T, Söderberg M, Wassélius J, Leander P (2015) Material
decom-position in dual-energy computed tomography separates high-Z
ele-ments from iodine, identifying potential contrast media tailored for dual
contrast medium examinations J Comput Assist Tomogr 39:975–980
30 Zhang LJ, Wang ZJ, Lu L et al (2011) Feasibility of
gadolinium-enhanced dual energy CT pulmonary angiography: a pilot study in
rabbits Int J Cardiovasc Imaging 27:1069–1080
31 Bongers M, Schabel C, Homann G et al (2015) Potential of gadolinium
as contrast material in second generation dual energy computed
tomog-raphy (DECT): an ex-vivo phantom study ECR 2015, C-1662
32 Strich G, Hagan PL, Gerber KH, Slutsky RA (1985) Tissue
distri-bution and magnetic resonance spin lattice relaxation effects of
gadolinium-DTPA Radiology 154:723–726
33 Cain PA, Ahl R, Hedstrom E et al (2009) Age and gender specific
normal values of left ventricular mass, volume and function for
gradient echo magnetic resonance imaging: a cross sectional study.
BMC Med Imaging 9:2
34 de la Grandmaison GL, Clairand I, Durigon M (2001) Organ weight
in 684 adult autopsies: new tables for a Caucasoid population.
Forensic Sci Int 119:149 –154
35 Geraghty EM, Boone JM, McGahan JP, Jain K (2004) Normal organ
volume assessment from abdominal CT Abdom Imaging 29:482 –490
36 Armstrong JD, Gluck EH, Crapo RO et al (1982) Lung tissue vol-ume estimated by simultaneous radiographic and helium dilution methods Thorax 37:676 –679
37 Johnson TR, Fink C, Schönberg SO, Reiser M (eds) (2011) Dual energy CT in clinical practice Medical radiology Springer-Verlag, Berlin
38 Hubbell JH, Seltzer SM (2004) Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients (version 1.4).
http://physics.nist.gov/xaamdi
39 van Hamersvelt RW, Willemink MJ, Takx RAP et al (2014) Cardiac valve calcifications on low-dose unenhanced ungated chest com-puted tomography: inter-observer and inter-examination reliability, agreement and variability Eur Radiol 24:1557–1564
40 Jones M, Dobson A, O’Brian S (2011) A graphical method for assessing agreement with the mean between multiple observers using continuous measures Int J Epidemiol 40:1308–1313
41 Halvorsen RA (2008) Which study when? Iodinated contrast-enhanced
CT versus gadolinium-enhanced MR imaging Radiology 249:9–15
42 Kanda T, Ishii K, Kawaguchi H et al (2014) High signal intensity in the dentate nucleus and globus pallidus on unenhanced T1-weighted MR images: relationship with increasing cumulative dose
of a gadolinium-based contrast material Radiology 270:834–841
43 McDonald RJ, McDonald JS, Kallmes DF et al (2015) Intracranial gadolinium deposition after contrast-enhanced MR imaging Radiology 275:772–782
44 Thomsen HS, Morcos SK, Almén T et al (2013) Nephrogenic systemic fibrosis and gadolinium-based contrast media: updated ESUR Contrast Medium Safety Committee guidelines Eur Radiol 23:307 –318
45 Leiner T, Kucharczyk W (2009) NSF prevention in clinical prac-tice: summary of recommendations and guidelines in the United States, Canada, and Europe J Magn Reson Imaging 30:1357 –1363
46 Murata N, Murata K, Gonzalez-Cuyar LF, Maravilla KR (2016) Gadolinium tissue deposition in brain and bone Magn Reson Imaging 34:1359 –1365
47 Radbruch A, Weberling LD, Kieslich PJ et al (2015) High-signal intensity in the dentate nucleus and globus pallidus on unenhanced T1-weighted images Invest Radiol 50:805 –810