Comparison of dosimeter response of TLD 100 and ionization chamber for high energy photon beams at KIRAN Karachi in Pakistan The Egyptian Journal of Radiology and Nuclear Medicine xxx (2017) xxx–xxx C[.]
Trang 1Original Article
Comparison of dosimeter response of TLD-100 and ionization chamber
for high energy photon beams at KIRAN Karachi in Pakistan
Muhammad Waqara,d,⇑, Asdar Ul-Haqb, Syed Bilald, M Masoodc
a Nuclear Medicine Oncology and Radiotherapy Institute Nawabshah (NORIN), Pakistan
b
Karachi Institute of Radiotherapy and Nuclear Medicine (KIRAN), Pakistan
c
Pakistan Institute of Nuclear Science and Technology (PINSTECH), Pakistan
d
Pakistan Institute of Engineering and Applied Sciences (PIEAS), Pakistan
a r t i c l e i n f o
Article history:
Received 15 October 2016
Accepted 21 January 2017
Available online xxxx
Keywords:
TLD
Dosimeter
Ionization chamber
Phantom
Relative variation
a b s t r a c t
This study was conducted by measuring point dose at different depths of water phantom by using two types of dosimeters (PTW N30013 ionization chamber and TLD-100 chips) at Karachi institute of radio-therapy and nuclear medicine (KIRAN), Pakistan Two different TLD chips were used (circular pallets and square chips) The main aim of this study was to compare the responses of two different types of dosime-ter irradiated with 6 MV X-rays using same paramedosime-ters Both types of dosimedosime-ter were irradiated with dif-ferent dose value ranging from 25 cGy to 500 cGy The deviation between difdif-ferent shapes of TLD and ionization chamber remained within 5% limit Maximum deviations of circular pallets reading from that
of ion chamber are 3.56% at 1.5 cm depth and 4.91% at 5 cm depth For square chips maximum deviation happened to be 4.38% at 1.5 cm depth and 4.23% at 5 cm depth This measurement shows that TLDs are reliable tool for dosimetry regardless of their shape or manufacturer and they can be used as re-validation tool for ion chamber dosimetry
Ó 2017 The Egyptian Society of Radiology and Nuclear Medicine Production and hosting by Elsevier This
is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/
4.0/)
1 Introduction
Inter comparison is an important activity to ensure the
consis-tency of radiation dosimetry [1] The ionization chamber is a
benchmark of dosimeter, and had been known to provide the most
accurate and reliable results [2] One difficulty with ionization
chamber is that the measured dose can be perturbed by volume
averaging over their relatively large active volume[3] Practically
there are several things to consider, such as the large volume of
active and effective point of measurements[4] They can’t be used
in anthropomorphic for in vivo measurements When ionization
chamber measurements are impractical, it can be replaced by
TLD, especially in in vivo dosimetry The small size of TLD allows
it to be inserted into an anthropomorphic or water phantom for
dosimetry TLDs can be used to measure point doses with greater
precision in volume as their active volume can be made very small
as compared to ionization chamber For fractional dose
measure-ment, the corrections for energy and dose response are taken into account [5] To achieve dose measurement precision, TLD imple-mentation program requires a rigorous annealing and response measurement protocol, and routine QA of the TLD reader and annealing oven temperature control[6]
When radiation is incident on a crystal of thermo luminescent material, it excites it This crystal can’t be de-excited itself and energy of radiation remains trapped If this crystal is heated up
to a certain temperature, it is de-excited and releases the trapped energy in the form photon of visible light This light is detected
by PMTs The output of PMTs is directly related to the light output
of TL crystal, which is itself proportional to the dose absorbed in the crystal by incident radiation If output of PMTs is calibrated against the absorbed dose, TLDs can be used to assess the dose absorbed in TL crystal Once TL dosimeter is read, it can be reused after a process called annealing which eliminates any residual imperfection in crystal[7,8] Thermo-luminescence properties of crystal can be explained theoretically by energy band theory of solids[9]
LiF is the most common crystal used for Thermo luminescence dosimetry because of its tissue equivalence (Zeff= 8.2 compared to 7.4 for tissue) and its energy independent response in the range of
100 keV–1.3 MeV [10,11] TLD-100 is LiF crystal doped with
http://dx.doi.org/10.1016/j.ejrnm.2017.01.012
0378-603X/Ó 2017 The Egyptian Society of Radiology and Nuclear Medicine Production and hosting by Elsevier.
This is an open access article under the CC BY-NC-ND license ( http://creativecommons.org/licenses/by-nc-nd/4.0/ ).
Peer review under responsibility of The Egyptian Society of Radiology and Nuclear
Medicine.
⇑ Corresponding author at: Senior Medical Physicist, Nuclear Medicine Oncology
and Radiotherapy Institute Nawabshah (NORIN), Pakistan.
E-mail address: phy_waqar@yahoo.com (M Waqar).
Contents lists available atScienceDirect
The Egyptian Journal of Radiology and Nuclear Medicine
j o u r n a l h o m e p a g e : w w w s c i e n c e d i r e c t c o m / l o c a t e / e j r n m
Trang 2Magnesium and Titanium and is usually used for dosimetry in
radiotherapy
The light output of PMTs is plotted against temperature or time
that is called glow curve[12] Various curves plotted in glow curve
shows the released electrons of different energy, which were
trapped in the crystal At room temperature TLD-100 has different
peaks corresponding to different electron energy level [13,14]
Electrons trapped at lower energy levels are prone to fading, which
is loss of trapped charges before readout Factors, which affect the
shape of glow curve are annealing, heating rate and its uniformity,
size and history of sample, threading instrument
TLD-100s are heated before irradiation first at 400°C for one
hour followed by 100°C for two hours or at 800 °C for 24 h (called
annealing) This slow heating relieves trapped electron remained in
lower peaks of glow curve by decreasing their trapping efficiency
These electrons can also be released by post irradiation annealing
by heating at 100°C for 10 min The magnitude of these peaks
rapidly decreases with time after irradiation so they can be
released at lower temperature after irradiation After removing
peaks at lower values of temperature, glow curve shows more
pre-dictable behavior[15]
The aim of present study is to compare the responses of TLD
dosimeters with the other most widely used dosimeters, ion
cham-ber Ion chambers are commonly used for dosimetry in
radiother-apy Present study focuses on a point that response of TLD
dosimeters is in well agreement with ion chambers and can be
used as a re-validation tool for ion chamber dosimetry
2 Materials and Methods
In this study, two different batches of TLDs (LiF-Mg-Ti) having
different shape, dimensions and manufactured by different
well-known companies (HARSHAW, USA & MTS, Poland) were used
First batch comprised of 125 square shape TLD chips (TLD-100)
having dimensions of 3.2 3.2 0.9 mm3, manufactured by M/s
HARSHAW, USA The second batch comprised of 125 circular
pel-lets (MTS-N) having the dimensions of 4.5 0.9 mm3
manufac-tured by M/s MTS, Poland
The ionization chamber (PTW N30013 Serial number 0114,
PTW-FREIBURG Corporation, Germany) with volume 0.6 cm3was
used in this study It is often used for absolute dosimetry
measure-ment in radiotherapy because it has flat response and better signal
to noise ratio For measurement of the charges produced by photon
in water and PMMA solid phantoms, CNMC11 electrometer (model
No 5232) was used with IC The ionization chamber was calibrated
from Secondary Standard Dosimetry Laboratory (SSDL) PINSTECH,
Islamabad
Linear Accelerator with of 6 MV fixed energy (Primus plus)
manufactured by Siemens Medical systems, USA) and
manually-operated TLD Reader system (Thermo-Scientific Model 3500) with
heating rate 10°C/s was utilized to carry out this study
Before use, TLDs were first grouped according to their Element
Correction Factor (ECC) 125 TLDs of each group are randomly
selected and their ECC were manually calculated HARSHAW
3500 single chip TLD reader was used for this purpose Around
140 TLDs with their ECC close or equal to unity were selected for
the measurement Among 140 TLDs, 70 were square chips and 70
were circular pellets
After grouping, TLDs were calibrated After each use, TLDs were
annealed in Thermo Lyne 47900 furnace, first at 400°C for one
hour and then at 100°C for two hours A time gap of almost 48 h
is given between exposure and reading the TLD to minimize the
fading Background or zero readings were rechecked after
anneal-ing, along with test light reading
A set of annealed TLD-100 were irradiated with field size
10 10 cm2, at 100 cm SSD and at 5 cm depth in a solid water phantom to a dose of 200 cGy with a 6 MV photon beam A calibra-tion factor was measured for each TLD All TLDs were irradiated repeatedly with 50 cGy, 100 cGy and 150 cGy to check the linearity and reproducibility of TLD response All the TLDs showed a linear response with this range of doses
The responses of TLD samples and ionization chamber were observed at SSD 100 cm with field size 10 10 cm2in water phan-tom at depths of 1.5 cm (Dmax) and 5 cm by varying the doses from
25 cGy to 500 cGy Experimental arrangements are shown inFig 1 Relative variation of measured TLD readings (DTLD) and Ioniza-tion chamber readings (DIC) are calculated using following formula [16]:
d ð%Þ ¼ DTLD DIC
DIC
3 Results and Discussion
In this section, the response of both batches of TLDs against irradiation was compared with the ion chamber response sepa-rately It was found that TLD response is in a good agreement with ion chamber readings
3.1 At the Depth of Maximum Dose (1.5 cm) Tables 1 and 2were the measured dose values of TLDs (DTLD) and IC (DIC) for X-ray beams with energy 6 MV measured at depth
of 1.5 cm Average relative variation for square TLD chips from that
of IC readings were calculated The maximum and minimum rela-tive variation was found to be 4.41% and 0.29% for the doses of
250 cGy & 500 cGy respectively This data is plotted inFig 2
Fig 1 Experimental arrangements.
Table 1 Responses of TLD (square chips) and Ion chamber dosimeter at the depth of 1.5 cm Dose (cGy) Square chips 1.5 cm
D TLD (cGy) D IC (cGy) Relative variation (%)
150 146.50 147.08 0.40
200 190.15 196.11 3.04
250 234.44 245.25 4.41
300 281.43 294.33 4.38
350 336.91 343.36 1.88
400 376.94 392.33 3.92
450 433.16 441.55 1.90
500 489.37 490.77 0.29
Trang 3For circular pallets, the maximum and minimum relative
varia-tion was found to be 3.56% and 0.25% at dose values of 500 cGy &
150 cGy respectively as tabulated inTable 2and plotted inFig 3
For 1.5 cm depth, relative variation for square chips and circular
pallets with respect to ionization chamber was observed to be
within ±4.5% and ±4% respectively
3.2 At Dose Depth (5 cm) Measurements were also performed at dose depth of 5 cm Dose values of TLD’s (DTLD) and IC (DIC) for X-ray beams with energy
6 MV measured at depths of 5 cm for the square and circular chips with ionization chamber are shown in Tables 3 and 4 Average
Table 2
Responses of TLD (circular pellets) and Ion chamber dosimeter at the depth of 1.5 cm.
Dose (cGy) Circular Pellets 1.5 cm
D TLD (cGy) D IC (cGy) Relative variation (%)
150 147.45 147.08 0.25
200 197.18 196.11 0.55
250 253.54 245.25 3.38
300 300.03 294.33 1.94
350 346.06 343.36 0.79
400 397.92 392.33 1.42
450 427.95 441.50 3.07
500 473.30 490.77 3.56
Fig 2 Comparison of TLD (square chips) and ion chamber response at the depth of 1.5 cm.
Fig 3 Comparison of TLD (circular chips) and ion chamber response at the depth of 1.5 cm.
Table 3 Responses of TLD (square chips) and Ion chamber dosimeter at the depth of 5 cm Dose (cGy) Square chips 5 cm
D TLD (cGy) D IC (cGy) Relative variation (%)
150 127.40 127.79 0.30
200 164.71 170.36 3.31
250 204.40 212.94 4.01
300 245.33 255.73 4.07
350 285.73 298.36 4.23
400 340.05 341.05 0.29
450 377.92 383.68 1.50
500 417.08 426.31 2.16
Trang 4relative variation for square TLD chips from that of IC readings
were calculated The maximum and minimum relative variation
was found to be 4.23% and 0.29% for the doses of 350 cGy &
400 cGy respectively The data is given inTable 3and plotted in
Fig 4 For 5 cm dose depth, the relative variations within ±4%
and ±5% for square chips and circular pellets respectively as shown
inFig 5 While comparing IC response with TLD’s response in water phantom at different depths, it was found that TLD is a suitable detector for quality assurance or re-validation purpose, especially for the energies of therapeutic range, provided that they have been properly calibrated The dose re-validation is very important because of the direct involvement of human beings Very limited and insufficient literature was available to support the procedure adopted in this section Banjade et al.[17]proposed that accuracy
of better than 5% of measured dose can be achieved He cited the results of Marshall et al for the measurement of depth dose with accuracy of 5.3% According to ICRU recommendation the variation
of 3–5% between the delivered and prescribed dose is allowed in radiotherapy [18] At 10 cm depth and energies of 6 MV and
18 MV, variation is shown to about 0.1–1.3%[16] At low energy
in mammography the variation of results between ion chamber and TLS is 4–8%[19] Present measurements have shown relative variation within ±4% and ±5% of ion chamber measurement at 1.5 cm & 5 cm respectively It is also suggested that more studies must be conducted, for comparison of these two dosimeters
Table 4
Responses of TLD (circular pellets) and Ion chamber dosimeter at the depth of 5 cm.
Dose (cGy) Circular Pellets 5 cm
D TLD (cGy) D IC (cGy) Relative variation (%)
150 122.59 127.79 4.07
200 163.63 170.36 3.95
250 219.28 212.94 2.98
300 259.92 255.73 1.64
350 298.46 298.36 0.03
400 352.40 341.05 3.33
450 395.27 384.67 2.76
500 438.37 426.31 2.83
Fig 4 Comparison of TLD (square chips) and Ion chamber response at the depth of 5 cm.
Fig 5 Comparison of TLD (circular chips) and ion chamber response at the depth of 5 cm.
Trang 5responses at specified depths (1.5 cm & 5 cm), to have a better
understanding
4 Conclusion
Ion chambers are primary radiation detectors for radiation
dosimetry in radiotherapy Their relatively large active volume
and their vulnerability to changes in atmospheric conditions,
sometimes, make their results unreliable In present study, TLDs
showed the results of within 5% deviation from those of ion
cham-ber which is in a good agreement The responses of both types of
TLD lie within acceptable limits regardless of shape and
manufac-turer It can be concluded that TLDs can be used as a re-validation
tool for ion chamber dosimetry
Conflict of interest
Authors have nothing to disclose
Acknowledgement
The authors are grateful to Mr Muhammad Shahban, NORIN
cancer hospital Nawabshah, Mr Shamim Haider and Tauseef
Reh-man, KIRAN cancer hospital Karachi for their assistance and help
References
[1] Zˇivanovic´ MZ, Lazarevic´ ÐR, Ciraj-Bjelac OF, Stankovic´ SJ, C´eklic´ SM, Karadzˇic´
KS Intercomparisons as an important element of quality assurance in
metrology of ionising radiation Nucl Technol Radiat Prot 2015;30:225–31
[2] Nalbant N Pre-treatment dose verification of IMRT using Gafchromic EBT3
film and 2D array J Nucl Med Radiat Ther 2014;2014
[3] Fitriandini A, Wibowo W, Pawiro S Comparison of dosimeter response:
ionization chamber, TLD, and Gafchromic EBT2 film in 3D-CRT, IMRT, and SBRT
techniques for lung cancer J Phys: Conf Ser 2016;012006.
[4] McEwen M, Kawrakow I, Ross C The effective point of measurement of ionization chambers and the build-up anomaly in MV X-ray beams Med Phys 2008;35:950–8
[5] Olko P Advantages and disadvantages of luminescence dosimetry Radiat Meas 2010;45:506–11
[6] Low DA, Moran JM, Dempsey JF, Dong L, Oldham M Dosimetry tools and techniques for IMRT Med Phys 2011;38:1313–38
[7] Pradhan A, Bakshi A Calibration of TLD badges for photons of energy above
6 MeV and dosimetric intricacies in high energy gamma ray fields encountered
in nuclear power plants Radiat Prot Dosim 2002;98:283–90 [8] Rosenwald J-C, Nahum A, Chavaudra J, Carlsson GA, Dance D, Bielajew A, et al Handbook of radiotherapy physics theory and practice; 2008.
[9] Siddique R, Uddin Z, Hussain M Dosimetric evaluation and verification of external beam 3-D treatment plans in humanoid phantom using thermoluminescent dosimeters (TLDs) J Basic Appl Sci 2012;8:690–5 [10] Cherry P, Duxbury A Practical radiotherapy: physics and equipment John Wiley & Sons; 2009
[11] Nette H, Onori S, Fattibene P, Regulla D, Wieser A Coordinated research efforts for establishing an in international radiotherapy dose intercomparison service based on the alanine/ESR system Appl Radiat Isot 1993;44:IN1–IN11 [12] Taylor G, Lilley E The analysis of thermoluminescent glow peaks in LiF (TLD-100) J Phys D: Appl Phys 1978;11:567
[13] Anacak Y, Arican Z, Bar-Deroma R, Tamir A, Kuten A Total skin electron irradiation: evaluation of dose uniformity throughout the skin surface Med Dosim 2003;28:31–4
[14] Delgado A, Ros JG Evolution of TLD-100 glow peaks IV and V at elevated ambient temperatures J Phys D: Appl Phys 1990;23:571
[15] Derreumaux S, Chavaudra J, Bridier A, Rossetti V, Dutreix A A European quality assurance network for radiotherapy: dose measurement procedure Phys Med Biol 1995;40:1191
[16] Swinnen A Quality assurance in radiotherapy: development and validation of
a mailed dosimetry procedure for external audits using a multipurpose phantom and in vivo dosimetry Katholieke Universiteit Leuven; 2005 [17] Banjade D, Raj TA, Ng B, Xavier S, Tajuddin A, Shukri A Entrance dose measurement: a simple and reliable technique Med Dosim 2003;28:73–8 [18] Andreo P, Burns DT, Hohlfeld K, Huq MS, Kanai T, Laitano F, et al Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water IAEA TRS 2000;398
[19] Stanton L, Day J, Brattelli S, Lightfoot D, Vince M, Stanton R Comparison of ion chamber and TLD dosimetry in mammography Med Phys 1981;8:792–8