1. Trang chủ
  2. » Thể loại khác

Encyclopedia of geology, five volume set, volume 1 5 (encyclopedia of geology series) ( PDFDrive ) 151

1 6 0

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề Encyclopedia of Geology, Five Volume Set, Volume 1-5
Trường học University of Geology
Chuyên ngành Geology
Thể loại Encyclopedia
Định dạng
Số trang 1
Dung lượng 82,08 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

For the SiLi detector, the FWHM at energy E is given by FWHME¼ ðFWHM20þ 21:1FEÞ0:5, where FWHM0 is the resolution at energy 0 and F is the Fano factor, which is a measure of the statisti

Trang 1

of the Mn Ka peak This measure is chosen because

readily available 55Fe is a source of this X-ray line

For the Si(Li) detector, the FWHM at energy E is

given by FWHME¼ ðFWHM20þ 21:1FEÞ0:5, where

FWHM0 is the resolution at energy 0 and F is the

Fano factor, which is a measure of the statistical

fluctuations in the ionization and charge collection

processes.Table 3lists the performance of a range of

X-ray spectrometers

Other energy-dispersive spectrometer technology

Ger-manium detectors have properties similar to those of

Si(Li) detectors and are preferred for use at higher

energies They are found on AEM, PIXE, and

syn-chrotron X-ray fluorescence (SXRF) instruments The

silicon drift detector (SDD) is based on

charge-coupled semiconductor technology and can provide

energy resolution similar to that of the monolithic

Si-crystal EDS, but at a count rate of 500 kHz

Further-more, an energy resolution of 140 eV can be achieved

at only 13C The detector area can be made as

large as 400 mm2 so that low currents can be used

for high count rates It is possible to count at more

than 1 MHz, but the resolution degrades as the count

rate increases This makes the detector unsuitable for

quantitative analysis but ideal for mapping of mineral

grains

X-Ray bolometry has been developed using

thin-film Ag microcalorimeters, transition edge sensors,

and superconducting quantum interference devices

Such detectors have energy resolutions down to 2 eV

and count rates of only 1 kHz In theory, arrays of

these millimetre-sized devices could be constructed

giving a high overall count rate The operating

tem-perature is 70–100 mK and it is possible to achieve

this using multistage Peltier cooling and an adiabatic

demagnetization refrigerator

Matrix Corrections

X-Ray intensities are measured in units of counts per second per nanoampere of beam current The weight percent concentration of an element in a sample, Csamp, is related to the characteristic X-ray intensity, Isamp, by the equation Csamp¼ Cstd(Isamp/

Istd)([MATRIX]samp/[MATRIX]std), where [MATRIX] denotes the effect of the chemical composition of the matrix on the X-ray intensity and ‘std’ refers to a standard of known composition

There are four approaches to matrix corrections:

1 Empirical methods assume that each element lin-early influences the X-ray intensity of each other element A table of coefficients, analysed element against matrix element, is drawn up using extra-polations from measurements of binary alloys and solid solution series These are known as alpha coefficients

2 The ZAF corrections separately compute the effects of atomic number (Z), absorption (A), and secondary fluorescence (F): ZAF ¼ R/S f(w)(1 þ g), where R is the back-scattering fraction and S is the X-ray generation factor due to stopping power; both of these are functions of atomic number The function of the mass attenuation coefficient, f(w), corrects for the absorption of the X-rays as they pass through the sample towards the detector The additional contribution when a matrix X-ray fluoresces an analysed element (Em>Ec,a) is represented by g

3 The f(rz) methods: f is defined as the ratio of the X-ray intensity from a thin layer, dz, of sample at a mass depth (rz) to the X-ray intensity of a similar layer isolated in space The f(rz) procedures inte-grate this X-ray intensity ratio function, corrected for multicomponent systems, from the surface to a

Table 3 X ray spectrometer devices

Crystal/device a Dispersion method b Resolution FWHM (eV) c

Max practical c ount rate (kHZ) Energy range (keV) Collection area (mm 2 )

a LiF, Lithium fluoride; PET, pentaerythritol; TAP, thallium acid phthalate; WSix, tungsten silicides; Ge, germanium; SDD, silicon drift detector.

b WDS, Wavelength dispersive spectrometry; EDS, energy dispersive spectrometry.

c FWHM, Full width half maximum.

112 ANALYTICAL METHODS/Mineral Analysis

Ngày đăng: 26/10/2022, 10:37

TÀI LIỆU CÙNG NGƯỜI DÙNG

TÀI LIỆU LIÊN QUAN