This is to distinguish Fourier transform infrared spectrometryfrom frustrated total internal reflection; FTIR is now an infrequently used term for 1 The etymology of the term spectrometry
Trang 2Fourier Transform Infrared Spectrometry
Second Edition
PETER R GRIFFITHS
University of IdahoMoscow, IdahoJAMES A de HASETH
University of GeorgiaAthens, Georgia
WILEY-INTERSCIENCE
A JOHN WILEY & SONS, INC., PUBLICATION
Trang 4Fourier Transform Infrared Spectrometry
Trang 6Fourier Transform Infrared Spectrometry
Second Edition
PETER R GRIFFITHS
University of IdahoMoscow, IdahoJAMES A de HASETH
University of GeorgiaAthens, Georgia
WILEY-INTERSCIENCE
A JOHN WILEY & SONS, INC., PUBLICATION
Trang 7No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or
by any means, electronic, mechanical, photocopying, recording, scanning, or otherwise, except as permitted under Section 107 or 108 of the 1976 United States Copyright Act, without either the prior written permission of the Publisher, or authorization through payment of the appropriate per-copy fee to the Copyright Clearance Center, Inc., 222 Rosewood Drive, Danvers, MA 01923, (978) 750-8400, fax (978) 750-4470, or on the web at www.copyright.com Requests to the Publisher for permission should be addressed to the Permissions Department, John Wiley & Sons, Inc., 111 River Street, Hoboken,
NJ 07030, (201) 748-6011, fax (201) 748-6008, or online at http://www.wiley.com/go/permission Limit of Liability/Disclaimer of Warranty: While the publisher and author have used their best efforts in preparing this book, they make no representations or warranties with respect to the accuracy or completeness of the contents of this book and specifically disclaim any implied warranties of merchantability or fitness for a particular purpose No warranty may be created or extended by sales representatives or written sales materials The advice and strategies contained herein may not be suitable for your situation You should consult with a professional where appropriate Neither the publisher nor author shall be liable for any loss of profit or any other commercial damages, including but not limited to special, incidental, consequential, or other damages.
For general information on our other products and services or for technical support, please contact our Customer Care Department within the United States at (800) 762-2974, outside the United States at (317) 572-3993 or fax (317) 572-4002.
Wiley also publishes its books in a variety of electronic formats Some content that appears in print may not be available in electronic formats For more information about Wiley products, visit our web site at www.wiley.com.
Wiley Bicentennial Logo: Richard J Pacifico
Library of Congress Cataloging-in-Publication Data:
10 9 8 7 6 5 4 3 2 1
Trang 81.4 Widths of Bands and Lines in Infrared Spectra 10
v
Trang 93.3.2 Digitization Noise 66
Trang 105.8 Lamellar Grating Interferometers 138
6.1 Infrared Radiation Sources for Transmission
7.2.1 Effect of Resolution and
Trang 118.4.1 Short-Term Performance 181
8.4.4 Effect of Sample Diameter and Thickness 186
9.12 General Guidelines for Calibration Data Sets 220
Trang 1213.3 Infrared Reflection–Absorption Spectrometry
13.3.1 Effect of Incidence Angle and Polarization 282
13.5.2 Liquid Sampling for Near-Infrared
Trang 1314.5.2 Hyperspectral Imaging with a
14.5.5 Applications of Hyperspectral Imaging 319
16.3 Applications of Mid-Infrared Diffuse
Trang 1417.3 Infrared Emission Spectra of
18.4.2 Surface-Enhanced Raman Spectroscopy 389
Trang 15Photoacoustic Phase 428
21.1 Dynamic Infrared Linear Dichroism Measured
21.2 DIRLD Spectrometry with a Step-Scan Fourier
21.4 DIRLD Spectrometry with a FT-IR Spectrometery
21.5 Other Sample Modulation Measurements
21.5.1 Liquid-Crystal Electroreorientation 458
Trang 1623.3.1 Introduction 491
23.4.2 Mobile-Phase Elimination Techniques
Trang 18The advances in the field of Fourier transform infrared (FT-IR) spectrometry in thepast 20 years have been quite remarkable FT-IR spectrometers are installed in justabout every analytical chemistry laboratory in the developed world Actually, wesometimes wonder why so many people still refer to these instruments as FT-IRspectrometers, or more colloquially simply as FTIRs, rather than simply as infraredspectrometers, since almost all mid-infrared spectra are measured with these instru-ments We note that scientists who use nuclear magnetic resonance, the other tech-nique that has been revolutionized by the introduction of Fourier transformtechniques, no longer talk about FT-NMR, as continuous-wave instruments (e.g.,grating monochromators) are a distant memory Nonetheless, practitioners of infra-red spectrometry seem to want to recall the era of grating monochromators, eventhough the vast majority has never seen one!
This book is the second edition of a volume that was published in 1986 In thepast 20 years, an enormous body of work has been published in which the key mea-surements have been made on FT-IR spectrometers When we started to write thisnew edition, it was not our intention to give a compendium of all these applications.Instead, we have tried to give a description of the theory and instrumentation of FT-
IR spectrometry as it stands today Even with this limitation, the material has taken
23 chapters to cover, and we know that a number of topics has been omitted We askall our many friends whose work is not referenced in this book for their understand-ing and forgiveness All we can say is that had we reviewed all the important andelegant experiments that have been done over the past 20 years, the book wouldhave rivaled the size of the 4000-page-long Handbook of Vibrational Spectroscopythat one of us recently coedited Instead, what we have tried to do is to provideusers of FT-IR spectrometers with a reasonably detailed description of how theirinstruments work and the types of experiments that can be performed even onthe less expensive instruments
At this point we should note that the way that infrared spectroscopy is appliedhas changed dramatically over the past 20 or 30 years Whereas infrared spectro-metry once played an important role in the structural elucidation of new organiccompounds, this task is now largely accomplished by NMR, mass spectrometryand x-ray crystallography Why, then, is the popularity of infrared spectrometry
at an all-time high? The answer is in part to be found in the versatility of this nique and in part in the relatively low cost of the instrumentation The number ofapplications for which a careful measurement of the infrared spectrum will yieldimportant qualitative, quantitative, and/or kinetic information is limited only by
tech-xv
Trang 19glossary at the end of the Handbook of Vibrational Spectroscopy, and we use hisrecommendations in a consistent manner The book is largely about the measure-ment of spectra: hence the title ‘‘Fourier Transform Infrared Spectrometry.’’1Throughout the book we have tried to use the term wavenumber when we refer
to the abscissa scale of a spectrum in cm1, using the term frequency only whenreferring to the modulation of a signal in hertz We note that curve fitting anddeconvolution are often (incorrectly) used interchangeably and we have tried touse the terms correctly We would also note that the verb from which (de)convolu-tion is derived is (de)convolve (as in revolve/revolution and evolve/evolution) Both
of us wish to demonstrate that the five years we each spent learning Latin in highschool was not misspent!
Measured or measurable parameters end in the suffix -ance (e.g., transmittance,absorbance, reflectance) A spectrum that is plotted with one of these parameters asthe y-axis can validly be referred to as a transmittance, absorbance, or reflectancespectrum; otherwise, it should be referred to as a transmission, absorption, orreflection spectrum We particularly note how reflection spectroscopy has falleninto this misuse Unfortunately, diffuse reflectance spectroscopy and attenuatedtotal reflectance (ATR) spectroscopy are now part of many spectroscopists’ lexicon.Pedagogy has held sway, however, and we have attempted to use the correct termi-nology throughout the book One of us (P.R.G) particularly regrets the poor usage
of the term diffuse reflectance spectroscopy in his early papers on this subject Heregrets even more that he coined the term DRIFT for this technique, as drift has allthe wrong connotations for any spectroscopic measurement Shortly after the firstpapers on DRIFT were published, Bob Hannah of PerkinElmer showed that diffusereflection infrared spectra could be measured easily on grating spectrometers andcoined the acronym DRUIDS (diffuse reflection using infrared dispersive spectro-metry!) We hope that neither Bob nor Peter is forced to live in acronym hell asresult of their transgressions on this planet!
We would also like to note the reason for the hyphen between FT and IRthroughout the book This is to distinguish Fourier transform infrared spectrometryfrom frustrated total internal reflection; FTIR is now an infrequently used term for
1
The etymology of the term spectrometry is clear, but the meaning of spectroscopy is less so as spectra are no longer measured with spectroscopes In this book we use spectrometry to mean the measurement of spectra and spectroscopy to mean the science of obtaining qualitative and quantitative information from spectra.
Trang 20ATR, but nevertheless, this usage was introduced before FT-IR spectrometersbecame popular.
Finally, we would like to thank the many people who have either supplied rial for the various chapters in this book or proofread the work in one of its severaldrafts In particular, we would like to thank Richard Jackson, Bruce Chase, LarryNafie, Rina Dukor, John Chalmers, Milan Milosevic, Neil Everall, and Roger Jones,
mate-as well mate-as the members of our research groups, for their comments We gratefullyacknowledge the patience and good grace of the six (count ‘em!) Wiley editors whohave tried to extract the manuscript from us Finally, our wives, Marie and Leslie,deserve our unending gratitude for putting up with us over the many years that ithas taken to assemble the material for this book
Trang 22Chapter 1
INTRODUCTION
TO VIBRATIONAL SPECTROSCOPY
1.1 INTRODUCTIONInfrared (IR) spectrometry has changed dramatically over the past 40 years Inthe 1960s, undergraduate chemistry majors would learn that the primary use ofinfrared spectrometry was for the structural elucidation of organic compounds
In many large research laboratories, however, the structure of complex molecules
is now usually found by a combination of techniques, including two-dimensionalnuclear magnetic resonance (NMR), x-ray diffraction, and mass spectrometry,with IR spectrometry playing a less dominant, although still important role.For example, U.S pharmaceutical companies must still submit IR spectra aspart of their application to the Food and Drug Administration as evidence ofthe putative chemical structure, and in polymer laboratories infrared spectrome-try is still used as the primary instrument for the determination of molecularstructure
This is not to imply that molecular structure of simple organic molecules not be determined by infrared spectroscopy In fact, the information that can bededuced from an infrared spectrum is complementary to that of other methods,and infrared spectroscopy provides valuable information that is unattainable byother methods, as is shown in the remainder of the book More important, how-ever, a plethora of other applications became available with the advent in 1969 ofthe first commercial mid-infrared Fourier transform spectrometer with better than
can-2 cm1 resolution These include quantitative analysis of complex mixtures, theinvestigation of dynamic systems, biological and biomedical spectroscopy, micro-spectroscopy and hyperspectral imaging, and the study of many types of inter-facial phenomena All of these applications (and many more) are described inthis book Furthermore, because of the development of such sampling techniques
Fourier Transform Infrared Spectrometry, Second Edition, by Peter R Griffiths and James A de Haseth Copyright # 2007 John Wiley & Sons, Inc.
1
Trang 23they are superseded by newer instruments with higher speed, sensitivity, specificity,
or resolution In his 1973 editorial, Laitinen used infrared spectrometry to illustrate
an instrument in its seventh age In fact, the technique was in its second childhood!Let us first consider why FT-IR spectrometers have assumed such a position ofdominance for the measurement of infrared spectra
Survey spectra in the mid-infrared region are often measured at a resolution of
4 cm1 When such spectra between 4000 and 400 cm1 are measured with aprism or grating monochromator, only one 4-cm1 resolution element in the3600-cm1-wide spectral range of interest is measured at any instant; the remaining
899 resolution elements are not Thus, the efficiency of the measurement is onlyabout 0.1% It was typical for survey scans to take several minutes to measure,whereas the measurement of archival-quality spectra (measured at 1 to 2 cm1reso-lution) often took at least 30 minutes
In FT-IR spectrometry, all the resolution elements are measured at all times ing the measurement (the multiplex or Fellgett’s advantage) In addition, moreradiation can be passed between the source and the detector for each resolution ele-ment (the throughput or Jacquinot’s advantage) These advantages are discussed inChapter 7 As a result, transmission, reflection, and even emission spectra can bemeasured significantly faster and with higher sensitivity than ever before
dur-In this book we demonstrate how FT-IR spectrometry can not only be used tomeasure infrared spectra of the type of samples that have classically been investi-gated by infrared spectrometers for decades (i.e., gases, liquids, and bulk and pow-dered solids in milligram quantities), but that interfacial species, microsamples, andtrace analytes can now be characterized routinely Measurement times have beenreduced from minutes to fractions of a second; in special cases, reactions takingplace in less than a microsecond can be followed The physical properties of mate-rials can be correlated to the molecular structure by vibrational spectroscopy betterthan by any other analytical technique It is probably true to say that during themore than three decades following Laitinen’s editorial, infrared spectroscopy hasentered and passed from its second childhood into its fifth age Because of theremarkable advances made in the performance of FT-IR spectrometers, infraredspectrometry has matured to the point that it is used for the solution of a variety
of problems from the research lab to the manufacturing floor, and sales of infraredspectrometers are at an all-time high
The increased popularity of infrared spectrometry and the commercial ity of instruments that are ‘‘so simple that a child can operate them’’ have led to the
Trang 24availabil-unexpected consequence that many operators of FT-IR spectrometers have receivedlittle or no formal training in vibrational spectroscopy To serve these new players inthe ‘‘FT-IR game’’ and to help give them a better appreciation of how the measure-ment of infrared spectra may be optimized, a brief introduction to the origin ofvibrational spectra of gases, liquids, and solids is given in the remainder of thischapter In the rest of the book, we show how FT-IR spectrometers work andhow to measure the most accurate and information-rich infrared spectra from awide variety of samples.
1.2 MOLECULAR VIBRATIONSInfrared spectra result from transitions between quantized vibrational energy states.Molecular vibrations can range from the simple coupled motion of the two atoms of
a diatomic molecule to the much more complex motion of each atom in a largepolyfunctional molecule Molecules with N atoms have 3N degrees of freedom,three of which represent translational motion in mutually perpendicular directions(the x, y, and z axes) and three represent rotational motion about the x, y, and z axes.The remaining 3N 6 degrees of freedom give the number of ways that the atoms
in a nonlinear molecule can vibrate (i.e., the number of vibrational modes).Each mode involves approximately harmonic displacements of the atoms fromtheir equilibrium positions; for each mode, i, all the atoms vibrate at a certain char-acteristic frequency, ni The potential energy, VðrÞ, of a harmonic oscillator isshown by the dashed line in Figure 1.1 as a function of the distance between theatoms, r For any mode in which the atoms vibrate with simple harmonic motion(i.e., obeying Hooke’s law), the vibrational energy states, Vi v, can be described
Figure 1.1 Potential energy of a diatomic molecule as a function of the atomic displacement during a vibration for a harmonic oscillator (dashed line) and an anharmonic oscillator (solid line).
Trang 25most common unit of length is the centimeter, in which case the wavenumber hasunits of cm1and is given the symbolen by many chemists and s by many physi-cists The energy difference for transitions between the ground stateðvi¼ 0Þ andthe first excited stateðvi¼ 1Þ of most vibrational modes corresponds to the energy
of radiation in the mid-infrared spectrum (400 to 4000 cm1)
The motion of the atoms during the vibration is usually described in terms of thenormal coordinate, Qi The molecule is promoted to the excited state only if its dipolemoment, m, changes during the vibration [i.e., provided that ðqm=qQiÞ 6¼ 0] Formolecules with certain elements of symmetry, some vibrational modes may be degen-erate, so that more than one mode has a given vibrational frequency whereas othersmay be completely forbidden Thus, because of degeneracy, the number of fundamen-tal absorption bands able to be observed is often less than 3N 6 Because rotation of
a linear molecule about the axis of the bond does not involve the displacement of any
of the atoms, one of the rotational degrees of freedom is lost and linear moleculeshave an additional vibrational mode Thus, the number of modes of a linear molecule
is 3N 5, so that a diatomic molecule ðN ¼ 2Þ has a single vibrational mode.The actual variation of the potential energy as a function of the displacement ofthe atoms from their equilibrium positions is shown as a solid line in Figure 1.1.From this curve it can be seen that Eq 1.1 is valid only for low values of the vibra-tional quantum number and is not valid whenvi is large In practice, Vivmust bedescribed using an anharmonic (Morse-type) potential function This behavior isshown in Figure 1.1 as a solid line, and the potential energy is given to a firstapproximation by the expression
Trang 26For many vibrational modes, only a few atoms have large displacements and therest of the molecule is almost stationary The frequency of such modes is charac-teristic of the specific functional group in which the motion is centered and is mini-mally affected by the nature of the other atoms in the molecule Thus, theobservation of spectral features in a certain region of the spectrum is often indica-tive of a specific chemical functional group in the molecule Extensive spectra/structure correlation tables (often known as Colthup charts) have been developed
to allow chemists to assign one or more absorption bands in a given infrared trum to the vibrational mode(s) associated with a certain functional group Thesetables may be found in many textbooks on the interpretation of infrared spectra.Other bands involve the significant motion of only a few atoms, yet their frequencyvaries from one molecule to another containing the particular functional group.These modes are useful to distinguish one molecule from another that containssimilar functional groups and hence are often known as fingerprint bands.Skeletal modes involve significant displacements of many of the atoms in the mole-cule These bands are rarely used to indicate the presence or absence of a specific func-tional group but again, may be useful to distinguish between structurally similarcompounds The vibrational frequency of skeletal modes is usually quite low As aresult, they absorb long-wavelength radiation that is often below the cutoff of manymid-infrared detectors The far-infrared region of the spectrum (10 to 400 cm1) israrely used for structural elucidation but contains useful information on the vibration
spec-of heavy atoms (especially for inorganic compounds) and/or weak bonds such ashydrogen bonds
Every molecule has slightly different vibrational modes from all other molecules(with the exception of enantiomers) Thus, the infrared spectrum of a given mole-cule is unique and can be used to identify that molecule Infrared spectra give farmore detailed information than simply allowing the presence or absence of certainfunctional groups to be recognized In the past, many chemists had a solid under-standing of how to interpret infrared spectra, but spectral interpretation is starting tobecome a lost art, in part because of the emergence of nuclear magnetic resonance,mass spectrometry, and x-ray diffraction for this purpose, which are easier to inter-pret Although molecular orbital programs are starting to permit infrared spectra ofquite complex molecules to be calculated, perfect matching of calculated and mea-sured spectra has yet to be achieved Today, computer-assisted comparison of thespectra of unknowns to a large number of reference spectra in a database (spectral,
or library, searching) has become a far more popular way than manual tion to find the structure of a molecule from its infrared spectrum Because of thesubtle differences between the spectra of many compounds, the result of a compu-terized spectral search should never be assumed to give the true identity of a com-pound without visual comparison by the operator between the best match and theactual spectrum of the unknown
interpreta-For most pure compounds, a sample thickness of only about 10 mm is needed toyield a mid-infrared spectrum for which the bands are neither saturated (maximumtransmittance less than 1%) nor so weak that they require ordinate expansion It isoften inconvenient and sometimes impossible to prepare such thin samples In these
Trang 27region) In the last two decades, NIR spectrometry has become of tremendousimportance, in large part because of the very weakness of these bands For samplesthat are between about 0.1 and 5 mm in thickness, the NIR spectra are often muchmore appropriate for quantitative, and sometimes even qualitative, analysis than thecorresponding mid-infrared spectra of these samples Furthermore, samples do nothave to be mounted in salt cells, sources are more intense, and NIR detectors aremore sensitive than mid-infrared detectors NIR spectra are not as easy to interpret
as mid-infrared spectra, but they are very amenable to multivariate statistical lysis of the type that is now becoming common throughout analytical chemistry.Indeed, many of these algorithms were originally developed specifically for theanalysis of NIR spectra
ana-1.3 VIBRATION–ROTATION SPECTROSCOPYOne of the greatest strengths of infrared spectrometry is that samples in all phases
of matter may be studied Infrared spectra of gases, liquids, and solids have ferent characteristics, and it is essential that these differences be understood ifspectra of materials in each state are to be measured optimally and to yieldthe greatest amount of information For example, the spectra of small molecules
dif-in the vapor phase show considerable fine structure because transitions betweenquantized rotational energy levels occur at the same time as vibrational transi-tions Similar features are rarely seen in the spectra of larger molecules in thevapor phase (because the individual rotational transitions are too close together
to be resolved) or any molecule in the liquid state (because collisions occur at
a greater rate than the rotational frequency) The full theory of vibration–rotationspectroscopy is quite complex, and a detailed exposition of this subject is beyondthe scope of this chapter, but a brief introduction to vibration–rotation spectro-scopy is given below
The simplest vibration–rotation spectra to interpret are those of diatomic cules The rotational energy levels of diatomic molecules are characterized by a sin-gle rotational quantum number, J If the molecule is assumed to be a rigid rotor(i.e., its bond length remains constant no matter how rapidly the molecule rotates),the rotational energy is given by
Trang 28B, called the rotational constant, is given by
Diatomic molecules, XY, have a single fundamental vibrational mode, of numberen0, which is infrared active only if X6¼ Y For any allowed vibrational tran-sition of a gaseous diatomic molecule, there must be a simultaneous rotationaltransition; that is,
Thus, the vibration–rotation spectrum of a rigid diatomic molecule consists of a ies of equally spaced lines above and belowen0 that correspond toJ ¼ þ1 and
ser-J ¼ 1, respectively The series of lines below en0 ðJ ¼ 1Þ is known as the
P branch of the band, while the lines aboveen0 ðJ ¼ þ1Þ are known as the Rbranch BecauseJ 6¼ 0, there is no absorption line at en0.1
In practice, molecules are not rigid rotors, and centrifugal forces cause the length
of the bond between X and Y to increase as the angular velocity of the rotating cule increases The effect of centrifugal distortion is to increase the moment of iner-tia, decreasing the rotational constant, B, at high J To a first approximation, theeffect of centrifugal distortion is taken care of by adding a second term to Eq 1.3:
where D is the centrifugal distortion constant Usually, 0.1< B < 10 cm1 and
D 104cm1 Because of the effect of centrifugal distortion, the spacing of thelines in the P branch increases as the distance fromen0 increases while that of thelines in the R branch decreases
Trang 29Figure 1.2 Infrared active vibration–rotation fundamental bands of carbon dioxide: (a) antisymmetric stretching mode (n 3 ) for which the selection rule is n3 ¼ 1 and J ¼ 1; (b) bending mode (n 2 ) for which the selection rules is n2 ¼ 1 and J ¼ 0, 1.
Trang 30without a simultaneous change in J is permitted:
Thus, there is a strong line in the spectrum, known as the Q branch, corresponding
toJ ¼ 0 The reason that the selection rules are different for these two modes isbecause different symmetry elements of the linear CO2molecule are lost duringthese two vibrations
Linear molecules have two equal principal moments of inertia, corresponding torotation about the center of mass about two mutually perpendicular axes, with thethird principal moment equal to zero Nonlinear molecules usually have three dif-ferent moments of inertia In this case, the vibration–rotation spectrum can be verycomplex, even for a simple molecule such as water The rotational fine structure ofthe HOH bending mode of water is shown in Figure 1.3
The two molecules whose vibration–rotation spectrum is shown in Figures 1.2and 1.3, CO2and H2O, are often encountered as interferences when mid-infraredspectra are measured (although the rotational lines in the spectrum of CO2are oftenunresolved when the spectrometer resolution is 4 cm1or poorer) In fact, it is goodpractice to eliminate all traces of these molecules in the beam path of an infraredspectrometer by purging the instrument with dry CO2-free air or pure nitrogen gas,
as the bands shown in Figures 1.2 and 1.3 will often be seen in the spectra As notedabove, because collisions occur at a greater rate than the rotational frequency ofmolecules in the liquid state, no rotational fine structure is seen
Figure 1.3 Vibration–rotation spectrum of the H O H bending mode of water vapor.
Trang 31Gaussian; that is, the absorbance at any wavenumberen is given by
AðenÞ ¼ A0exp 4 ðln 2Þðen en0Þ2
g2 D
gD
en ¼ 7:16 107
ffiffiffiffiffiTM
r
ð1:12Þ
For a line in the HOH bending mode of water (M ¼ 18 g mol1) at 1500 cm1,the Doppler width at room temperature (298 K) is about 0.0044 cm1 Thus, aninstrument with very high resolution is needed before Doppler-broadened spectracan be measured accurately
As the total pressure of the gas is raised above 1 torr, the mechanism of line ening becomes more dominated by the effect of intermolecular collisions than by theDoppler effect The shape of lines in collision-broadened spectra is Lorentzian:
broad-AðenÞ ¼ A0
g2 C
g2
where gC, the FWHH of the collision-broadened line, is directly proportional to thepressure of the gas and increases with the polarity of each component There are sev-eral mechanisms of collision broadening, each of which leads to a slightly differentvariation of the broadening coefficient with temperature, from 1=pffiffiffiffiT
for hard-spherecollisions to 1/T for dipole–dipole interactions For many molecules in air at ambienttemperature, the collision-broadening coefficient is between 0.1 and 0.2 cm1 atm1
Trang 32Hence for mixtures of an analyte with helium, nitrogen, or air at atmospheric pressure,
gC of each of the rotational lines is usually between 0.1 and 0.2 cm1
For molecules at pressures between about 1 and 100 torr, the line width is mined by both Doppler and collision broadening In this case, the shape is given by
deter-a convolution of the shdeter-apes given by Eqs 1.10 deter-and 1.13; such deter-a shdeter-ape is known deter-as deter-aVoigt profile The higher the pressure of the gas, the greater the contribution of col-lision broadening to the Voigt profile Voigt profiles cannot be expressed analyti-cally, as they result from the convolution of a Lorentzian and Gaussian shape,but the FWHH is given to a good approximation by
gV ¼gC
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
gC2
a resolution much higher than 4 cm1to measure an easily recognizable spectrum,although to obtain a spectrum with minimal distortion of the rotational contour, asomewhat higher resolution should be used
1.4.2 Spectra of Condensed-Phase SamplesSince there is no rotational fine structure in the infrared spectra of liquids, theirspectra are much simpler than those of gases To a good approximation, the shape
of bands in the infrared spectra of liquids is Lorentzian (see Eq 1.13) In practice,the far wings of bands in the spectra of liquids die out somewhat faster than would
be given by Eq 1.13 To model the behavior of bands in the spectra of liquids,bands are sometimes expressed as the sum of Lorentzian and Gaussian bands:
AðenÞ ¼ ð1 aÞA0
g2 L
g2
Lþ ðen en0Þ2þ a A0exp ðln 2Þðen en0Þ2
g2 G
ð1:15Þ
where gLis the HWHH of the Lorentzian component, gGthe HWHH of the Gaussiancomponent (usually assumed to be the same), and a the Gaussian fraction For manyliquids, a< 0.1 Although complicated, this form is at least analytical, unlike theVoigt profile An even better description of bands in the spectra of liquids can bemade with the use of the classical damped harmonic oscillator model, but this isbeyond the scope of this chapter
The actual widths of absorption bands in the mid-infrared spectra of liquids andsolutions depend strongly on the rigidity of the part of the molecule where the
Trang 33are rarely smaller than 4 cm , it is common for survey spectra of liquids to bemeasured at a resolution of 4 or 8 cm1, whereas high-quality spectra for archivalpurposes are usually measured at a resolution of 1 or 2 cm1 Of course, measure-ment of a spectrum with a certain signal-to-noise ratio (SNR) takes far longer at aresolution of 1 cm1than the corresponding measurement at a resolution of 8 cm1.The ‘‘trading rules’’ among resolution, measurement time, and SNR are covered insome detail in Chapter 7.
Few detailed investigations into the shapes and widths of bands in the spectra ofsolids have been made, but the bands of solids are usually narrower than the corre-sponding bands of the same molecule in solution because of the restricted motion ofthe functional groups in the molecules The shapes of mid-infrared absorptionbands of solid compounds are often represented by Eq 1.15, with no limitation
as to the value of the Gaussian fraction, a
Because bands in near-infrared spectra are overtones or combinations of mentals, the widths of these bands are typically greater than the widths of bandsfrom which they are derived For example, the FWHH of the first overtone of aCH stretching band is to a first approximation twice that of the correspondingfundamental As a result, most NIR spectra of liquids are measured at significantlylower resolution than is the corresponding mid-infrared spectrum
funda-1.5 QUANTITATIVE CONSIDERATIONS
1.5.1 Beer’s LawThe Bouguer–Lambert–Beer law (usually called simply Beer’s law) is the funda-mental law of quantitative spectroscopy and is derived in all elementary textbooks
on instrumental analysis The transmittance of any sample at wavenumber en isgiven by the ratio of the radiant power emerging from the rear face of the sample
at that wavenumber IðenÞ to the power of the radiation at the front face of the ple, I0ðenÞ The transmittance of a pure sample of thickness b (cm) at wavenumber en
sam-is given by Beer’s law as
TðenÞ ¼ IðenÞ
Trang 34where aðenÞ is the linear absorption coefficient (cm1) aten The absorbance of thesample aten, AðenÞ, is given by the base 10 logarithm of 1=TðenÞ:
AðenÞ ¼ log10
1TðenÞ¼
1
ð1=ln 10ÞaðenÞ is the absorptivity at en, aðenÞ
If the sample is a mixture, the absorbance of each component, i, at concentration,
ci, is given by Beer’s law as
log10 1
where aiðenÞ has the units of (concentration pathlength)1 For N-component tures where more than one component absorbs aten, the total absorbance is givenby
trans-As we will see in Chapter 2, FT-IR spectrometers measure interferograms, fromwhich a single-beam spectrum is calculated The intensity of the single-beamspectrum, BðenÞ, at any wavenumber en is proportional to the power of the radiationreaching the detector Thus, to measure the absorbance spectrum of a sample, theratio of the single-beam spectra of the sample and background is first calculated (toyield the transmittance spectrum), which is then converted to absorbance as shown
in Eq 1.18 It should always be remembered that as the first step of many of themore popular operations in FT-IR spectrometry, including spectral subtraction, mul-ticomponent analysis, and spectral searching, the measured spectrum should always
be converted to absorbance
Second, the effect of reflection loss at the windows of the cell has been neglected
in the treatment described above The refractive index, n, of most organic samplesand windows is about 1.5, so that the reflectance of the front surface is about 4%(see Section 13.2.2) If Beer’s law is to be applied accurately, the apparent absor-bance caused by reflection loss (0.018 absorbance unit for windows with n¼ 1:5)should first be subtracted from the measured absorbance spectrum
Trang 35For any material, nðenÞ is determined by Snell’s law A few materials have nosignificant absorption in the mid- and near infrared Those materials with lowrefractive index ð1:45 n 1:6Þ are useful as windows, whereas those withhigh refractive indexð2:4 n 4:0Þ are frequently used as internal reflection ele-ments (see Chapter 15) For organic and inorganic molecules whose spectra exhibittypical absorption bands, the refractive index changes across the absorption band Atypical refractive index spectrum has the appearance shown in Figure 1.4a This
Figure 1.4 (a) Refractive index and (b) absorption index spectra of poly(methyl methacrylate).
Trang 36type of variation in the refractive index across absorption bands is known as alous dispersion Outside those regions of anomalous dispersion, n is fairly constantfrom the visible through the mid-infrared spectrum.
anom-Like the real refractive index, the imaginary refractive index is also a less quantity For pure materials, kðenÞ is given by
where aðenÞ is the linear absorption coefficient defined in Eq 1.16 A typical tion index spectrum looks like an absorbance spectrum (see Figure 1.4b) Whereasthe transmission spectrum of samples is largely determined by the value of kðenÞ foreach band, both of the optical constants nðenÞ and kðenÞ control the reflection of sam-ples, as discussed in Chapter 13
absorp-1.6 POLARIZED RADIATION
If a beam of unpolarized electromagnetic radiation is transmitted in thez direction,the amplitudes of the components of the sinusoidally varying electric field in the xand y planes are identical When the beam is passed through a polarizer, the com-ponent of the electric field in one plane is transmitted, as described in more detail inChapter 12 For unoriented samples such as all gases and liquids and isotropicsolids, the absorbance of all bands in the spectrum is independent of the orientation
of the polarizer If the molecules in a certain sample are preferentially oriented in agiven direction, however, the component of the dynamic dipole moment derivative
of each vibrational mode, dm=dQ, in the direction that the radiation is polarized willchange as the polarizer is rotated
Since the largest intensity of each band is observed when the beam is polarized
in the direction for which the change in dipole moment for that vibrational mode isgreatest, important information on the orientation of samples can be derived byinstalling a polarizer in the beam For example, one of the more important applica-tions of the use of polarized light is the measurement of the orientation of the chains
in drawn or extruded polymers A more esoteric application is the estimation of theangle at which surfactants are adsorbed on the surface of water
Several different types of reflection spectroscopy yield more information whenthe radiation is polarized For example, very thin films of molecules adsorbed onthe surface of metals only absorb radiation polarized parallel to the plane of inci-dence The reflection of light from bulk samples depends on the polarization of thelight with respect to the plane of the sample The effective depth into a sample thatcan be sensed by internal reflection spectroscopy is also different for radiationpolarized perpendicular and parallel to the surface Polarized radiation may even
be used to eliminate interference fringes from the spectra of thin polymer films
It is hoped that these few examples will give the less experienced reader an cation as to why FT-IR spectroscopy is even more popular today than when it wasused primarily as a tool for structural elucidation
Trang 37the incident photon Since the energies of the incident and scattered photons areidentical, this process is a form of elastic scattering often referred to as Rayleighscattering A small fraction of the incident photons drop back to the first excitedvibrational state of the ith vibrational mode of the molecule, so that the energy
of the scattered photon is hcðen0eniÞ i.e., it will be observed at a wavenumber
of ðen0eniÞ, whereeni is the wavenumber of the ith vibrational mode Since theenergies of the incident and scattered photons are different, the scattering is inelas-tic This process is known as Stokes Raman scattering Since a given molecule hasmany different vibrational modes, measurement of the spectrum from en0 toðen0 4000Þ cm1allows all Raman-active vibrational bands to be measured.Not all molecules are in the ground vibrational state A few molecules are in thefirst excited state of each vibrational mode,vi¼ 1 Neglecting the effect of degen-eracy, the Boltzmann population of molecules in the first vibrational state of the ithmode is given by
N1
where N1and N0are the numbers of molecules in the first excited state and groundstate of the ith vibrational mode, respectively When illuminated by a laser, a fewmolecules in an excited vibrational state can be promoted to a virtual level,hcðen0þeniÞ, and then return to the ground state, resulting in Raman bands at wave-numbers above that of the laser This process, known as anti-Stokes scattering, ismuch weaker than Stokes scattering for bands above about 500 cm1 above en0(where the Boltzmann population is less than 1%)
The intensity of bands in the Raman spectrum of a compound are governed bythe change in polarizability, a, that occurs during the vibration The intensity of anyband in the Raman spectrum is given by the following expression:
Trang 38dependent on the optical geometry, collection efficiency, detector sensitivity, andamplification.
Before the mid-1980s, Raman spectroscopy was often considered to be lessdesirable to infrared absorption spectroscopy for two reasons First, only one inbetween 108and 1010of the incident photons undergoes Raman scattering Thus,until about 1985, Raman spectrometry was considered to be a relatively insensitiveand/or time-consuming technique However, the advent of multiplex (usually, Four-ier transform) and multichannel (monochromators with array detectors) techniqueshas greatly increased the sensitivity of modern Raman spectrometry Of greater dif-ficulty today is the fact that many compounds fluoresce when illuminated by visiblelasers, such as by the radiation from an Arþlaser at 488 nm For some moleculesthe quantum efficiency for fluorescence with 488-nm radiation can approach 100%,and it is often greater than 0.001% Even in the latter case, the radiation emitted byluminescence far exceeds the intensity of Raman-scattered radiation Similarly,fluorescence from a trace molecule with a quantum efficiency approaching 100%can also swamp the Raman signal
To ameliorate the problem of fluorescence, near-infrared lasers can be used toilluminate the sample By the use of NIR radiation, the wavelengths used to exciteRaman spectra are long enough that most compounds no longer fluoresce Twotypes of NIR lasers can be used to measure Raman spectra The first is the diodelaser, with the most popular emitting at785 or 840 nm (12,740 or 11,900 cm1,respectively) Raman spectra generated with NIR diode lasers can be measuredusing silicon charge-coupled device (CCD) array detectors, which cut on at about
9500 cm1, limiting the Raman spectrum to about 3240 and 2500 cm1, depending
on whether a 785- or 840-nm laser is being used
The other popular NIR laser is the Nd : YAG laser, which emits at 1064 nmðen0 9400 cm1Þ With this laser, the problem changes from fluorescence to detec-tion Consider a Raman band due to a CH stretching mode at 2950 cm1 This bandmust be measured at (94002950) cm1(i.e., 6450 cm1), which is well below thewavenumber at which silicon charge-coupled device (CCD) array detectors cut off.Detectors operating in this region of the spectrum are significantly less sensitive thanCCDs Furthermore, it can be seen from Eq 1.23 that IRaman is proportional toðen0eniÞ4
Thus, a Raman band at 2950 cm1measured with a 1064-nm Nd : YAGlaser would be about 55 times weaker than the same band measured with the 488-nmline of an Ar+laser of equal power Since CCD array detectors do not respond at
6450 cm1, it is not surprising that Fourier transform techniques have been invokedfor the measurement of weak Raman signals at such long wavelengths
There are many reasons why scientists want to measure the Raman spectra of pounds First, many bands that are weak in the infrared spectrum are among the stron-gest bands in the Raman spectrum For example, the SS and CC stretching bandsare often so weak as to be essentially unrecognizable in the IR spectrum but stick outlike the proverbial sore thumb in a Raman spectrum Second, some Raman bands arefound at very characteristic frequencies For instance, monosubstituted aromaticcompounds, together with 1,3-disubstituted and 1,3,5-trisubstituted aromatics, have
com-a very intense bcom-and com-at 1000 cm1 This band, along with the presence or absence
Trang 39ond childhood yet is debatable Nonetheless, Raman spectrometry is still a vitalweapon in a vibrational spectroscopist’s arsenal.
1.8 SUMMARY
In Chapters 2 to 8 we describe the theory and instrumentation needed for an ciation of the way that Fourier transform infrared and Raman spectra are measuredtoday The sampling techniques for and applications of FT-Raman spectrometry aredescribed in Chapter 18 The remaining chapters cover the techniques and applica-tions of absorption, reflection, emission, and photoacoustic spectrometry in themid- and near-infrared spectral regions
appre-REFERENCE
1 H A Laitinen, Anal Chem 45, 2305 (1973)
Trang 40Chapter 2
THEORETICAL BACKGROUND
2.1 MICHELSON INTERFEROMETERThe design of many interferometers used for infrared spectrometry today is based
on that of the two-beam interferometer originally designed by Michelson in 1891[1,2] Many other two-beam interferometers have subsequently been designed thatmay be more useful than the Michelson interferometer for certain specific applica-tions Nevertheless, the theory behind all scanning two-beam interferometers issimilar, and the general theory of interferometry is most readily understood by firstacquiring an understanding of the way in which a simple Michelson interferometercan be used for the measurement of infrared spectra
The Michelson interferometer is a device that can divide a beam of radiation intotwo paths and then recombine the two beams after a path difference has been intro-duced A condition is thereby created under which interference between the beamscan occur The variation of intensity of the beam emerging from the interferometer
is measured as a function of path difference by a detector The simplest form of theMichelson interferometer is shown in Figure 2.1 It consists of two mutually per-pendicular plane mirrors, one of which can move along an axis that is perpendicular
to its plane
Bisecting the fixed mirror and the movable mirror is a beamsplitter, where a limated beam of radiation from an external source can be partially reflected to thefixed mirror (at point F for the median ray) and partially transmitted to the movablemirror (at point M) When the beams return to the beamsplitter, they interfere andare again partially reflected and partially transmitted Because of the effect of inter-ference, the intensity of each beam passing to the detector and returning to thesource depends on the difference in path of the beams in the two arms of the inter-ferometer The variation in the intensity of the beams passing to the detector andreturning to the source as a function of the path difference ultimately yields thespectral information in a Fourier transform spectrometer
col-Fourier Transform Infrared Spectrometry, Second Edition, by Peter R Griffiths and James A de Haseth Copyright # 2007 John Wiley & Sons, Inc.
19